4.7 Article

Detection of SARS-CoV-2 Using Reverse Transcription Helicase Dependent Amplification and Reverse Transcription Loop-Mediated Amplification Combined with Lateral Flow Assay

Journal

BIOMEDICINES
Volume 10, Issue 9, Pages -

Publisher

MDPI
DOI: 10.3390/biomedicines10092329

Keywords

RT-HDA; RT-LAMP; diagnostics; point-of-care; SARS-CoV-2; isothermal amplification; LFA

Funding

  1. National Centre for Research and Development [TANGO-IV-A/0015/2019, POIR.01.01.01-00-0638/20]

Ask authors/readers for more resources

This study presents the application of two nucleic acid isothermal amplification methods combined with lateral flow assay for the rapid detection of SARS-CoV-2. The RT-HDA combined with LFA method is found to be more sensitive and practical.
Rapid and accurate detection and identification of pathogens in clinical samples is essential for all infection diseases. However, in the case of epidemics, it plays a key role not only in the implementation of effective therapy but also in limiting the spread of the epidemic. In this study, we present the application of two nucleic acid isothermal amplification methods-reverse transcription helicase dependent amplification (RT-HDA) and reverse transcription loop-mediated amplification (RT-LAMP)-combined with lateral flow assay as the tools for the rapid detection of SARS-CoV-2, the etiological agent of COVID-19, which caused the ongoing global pandemic. In order to optimize the RT-had, the LOD was 3 genome copies per reaction for amplification conducted for 10-20 min, whereas for RT-LAMP, the LOD was 30-300 genome copies per reaction for a reaction conducted for 40 min. No false-positive results were detected for RT-HDA conducted for 10 to 90 min, but false-positive results occurred when RT-LAMP was conducted for longer than 40 min. We concluded that RT-HDA combined with LFA is more sensitive than RT-LAMP, and it is a good alternative for the development of point-of-care tests for SARS-CoV-2 detection as this method is simple, inexpensive, practical, and does not require qualified personnel to perform the test and interpret its results.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available