Influence of NiTi Wire Diameter on Cyclic and Torsional Fatigue Resistance of Different Heat-Treated Endodontic Instruments

Journal

MATERIALS
Volume 15, Issue 19, Pages -

Publisher

MDPI
DOI: 10.3390/ma15196568

Keywords

cyclic fatigue; endodontics; heat treatment; NiTi wire diameter; torsional fatigue; 2Shape

Ask authors/readers for more resources

In this study, we compared the mechanical properties of 2Shape mini TS2 instruments made from different sizes and thermally treated nickel-titanium wires. The results showed that files made from controlled memory wire had higher cyclic fatigue resistance regardless of the wire size, and testing at body temperature had a negative impact on the performance of all files. Additionally, the 1.0mm diameter T-wire instruments exhibited higher cyclic fatigue resistance compared to the 1.2mm diameter instruments.
We compared the mechanical properties of 2Shape mini TS2 (Micro-Mega, Besancon, France) obtained from 1.0 diameter nickel-titanium (NiTi) wires and 2Shape TS2 from 1.2 diameter nickel-titanium (NiTi) wires differently thermally treated at room and body temperature. We used 120 NiTi TS2 1.0 and TS2 1.2 files made from controlled memory (CM) wire and T-wire (n = 10). Cyclic fatigue resistance was tested by recording the number of cycles to fracture (NCF) at room and body temperatures using a customized testing device. Maximum torque and angle of rotation at failure were recorded, according to ISO 3630-1. Data were analyzed by a two-way ANOVA (p < 0.05). The CM-wire files had significantly higher NCFs at both temperatures, independent of wire dimensions. Testing at body temperature negatively affected cyclic fatigue of all files. The 1.0-mm diameter T-wire instruments showed higher NCF than the 1.2-mm diameter, whereas no significant differences emerged between the two CM wires at either temperature. The maximum torque was not significantly different across files. The TS2 CM-wire files showed significantly higher angular rotation to fracture than T-wire files. The TS2 CM-wire prototypes showed higher cyclic fatigue resistance than T-wire prototypes, regardless of wire size, exhibiting suitable torsional properties. Torsional behavior appears to not be affected by NiTi wire size.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available