4.6 Article

Comparison of the performance of Aptima HIV-1 Quant Dx Assay with Abbott RealTime HIV Assay for viral load monitoring using plasma and Dried Blood Spots collected in Kenya

Journal

PLOS ONE
Volume 17, Issue 8, Pages -

Publisher

PUBLIC LIBRARY SCIENCE
DOI: 10.1371/journal.pone.0269838

Keywords

-

Funding

  1. Hologic, Inc

Ask authors/readers for more resources

This study evaluated the performance characteristics of the Hologic Aptima HIV-1 Quant Dx Assay for viral load monitoring using dried blood spots (DBS) and plasma samples in Kenya. The study also determined the stability of HIV-1 RNA in DBS. The results showed that the Aptima assay had high agreement with the Abbott RealTime HIV-1 Assay and demonstrated accurate VL monitoring in both DBS and plasma samples.
Introduction HIV-1 viral Load (VL) testing is recommended for the monitoring of antiretroviral treatment. Dried Blood Spots (DBS) are an effective sample type in resource limited settings, where safe phlebotomy and reliable shipping are hard to guarantee. In HIV high burden countries, high throughput assays can improve access to testing services. The Hologic Aptima HIV-1 Quant Dx Assay (Aptima Assay) is a high throughput assay that runs on the CE-IVD approved Panther platform. The objectives of this study were to assess the performance characteristics of Aptima for VL monitoring using plasma and venous DBS specimens and to determine the stability of HIV-1 RNA in DBS. Materials and methods This was a cross-sectional study of 2227 HIV infected adults visiting health facilities in Nairobi and Busia, Kenya. Each provided a venous blood sample; plasma was prepared from 1312 samples while paired DBS samples and plasma were prepared from the remaining 915 samples. The agreement between the Aptima assay and the Abbott RealTime HIV-1 Assay (Abbott RT) was analysed by comparing the HIV-1 VL in both assays at the medical decision point of 1000 copies/mL. To assess stability of HIV-1 RNA in DBS, VL in DBS spotted on day 0 were compared with that from the same DBS card after 21 days of storage at room temperature. Results Overall, 436 plasma samples had quantifiable results in both Aptima and Abbott RT. The agreement between the two assays at 1000 copies/mL was 97.48% with a Pearson' correlation coefficient (r) of 0.9589 and gave a mean bias of 0.33 log copies/mL on Bland-Altman analysis. For fresh DBS, the agreement in both assays was 94.64% at 1000 copies/mL, with an r of 0.8692 and a mean bias of 0.35 log copies/mL. The overall agreement between DBS tested in Aptima on day 0 versus day 21 was 95.71%, with a mean bias of -0.154. Conclusion The Aptima HIV-1 Quant Dx assay is an accurate test for VL monitoring of HIV-1 using DBS and plasma sample types in Kenya.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available