4.3 Article

The performance of a home-made 300 kV AMS system at CIAE

Publisher

ELSEVIER
DOI: 10.1016/j.nimb.2022.07.008

Keywords

Accelerator mass spectrometry; 300 kV; Home-made; I-129; Pu-239

Funding

  1. National Natural Science Foundation of China [11875326]
  2. Continuous Basic Scientific Research Project [WDJC-2019-04]

Ask authors/readers for more resources

The China Institute of Atomic Energy has successfully developed a homemade 300 kV AMS system for heavy isotopes measurement, achieving good transmission efficiency and uncertainty in the measurement of I-129 and Pu-239.
In order to develop a AMS system for heavy isotopes measurement at China Institute of Atomic Energy (CIAE), a home-made 300 kV AMS system has been established. The measurement methods for I-129 and Pu-239 have been developed with the system. The transmission efficiency of 65 % for I-129 and 25 % for Uranium have been obtained with this system. The I-129 measurements with uncertainty of 2.5 % and sensitivity of I-129/I-127 = 2 x 10(-14) are achieved. The uncertainty of 5 % and detection limitation of 0.5 fg are obtained for Pu-239 measurements.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.3
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available