4.3 Article

Method development for on-site monitoring of volatile organic compounds via portable TD-GC-MS: evaluation of the analytical performances of HAPSITE® ER instrumentation and thermal desorption sampling media

Publisher

TAYLOR & FRANCIS LTD
DOI: 10.1080/03067319.2022.2121163

Keywords

HAPSITE (R) ER; TD-GC-MS; gas chromatography; thermal desorption

Funding

  1. Oak Ridge Institute for Science and Education (ORISE)

Ask authors/readers for more resources

Determining worker exposure to hazardous volatile organic compounds (VOCs) in air exceeding regulatory limits is important for occupational health assessments. This study aimed to develop a reliable method for on-site assessments using a portable instrument. Various parameters were determined to optimize the analysis, and calibration curves were generated. The method was successfully applied to chloroform, benzene, trichloroethylene, and heptane, but some sorbents showed limitations. Further research is currently being conducted to improve the method.
Determining worker exposure to hazardous volatile organic compounds (VOCs) in air at levels exceeding the Permissible Exposure Limits and Recommended Exposure Limits established by the U.S. federal agencies of Occupational Safety and Health Administration (OSHA) and the National Institute for Occupational Safety and Health (NIOSH), respectively, will continue to be an important part of environmental and occupational health risk assessments. The purpose of this work was to develop a reliable analytical method for rapid and on-site assessments of occupational VOC exposures using field-capable thermal desorption-gas chromatography-mass spectrometry (TD-GC-MS) instrumentation (i.e. the HAPSITE (R) ER). The experiments involved in this study included determining TD-GC-MS parameters suitable for efficient analyte separation and quantitation on the HAPSITE (TM) ER, determinations of analyte mass loadings that cause mass spectrometer detector saturations, generation of calibration curves, estimations of the limits of detection (LODs) and quantification (LOQs), as well as desorption efficiency and relative response factor repeatability. The LODs using Carbopack (TM) B and Tenax (R) TA sampling media were estimated and ranged from 0.2-1.9 ng and 0.045-0.3 ng, respectively. The LOQs using Carbopacle (TM) B and Tenax TA sampling media were estimated and ranged from 1.0-6.3 ng and 0.2-1.1 ng, respectively. We have developed a reliable analytical method for chloroform, benzene, trichloroethylene, and heptane using field-portable HAPSITE (R) ER instrumentation and Tenax (R) TA sorbent media. Reliable and accurate methods were developed for chloroform and trichloroethylene using Carbopack (TM) B sorbent media, however, this particular sorbent hadlow desorption efficiency and insufficient repeatability in relative response factors for many analytes. Our current and ongoing work in determining the uptake rates for analytes on Tenax (R) TA sorbent media will make the methods described herein applicable for on-site occupational VOC exposure assessments of chloroform, benzene, trichloroethylene, and heptane using either passive or active air sampling techniques.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.3
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available