4.6 Article

Clinical evaluation of the OC-Sensor Pledia calprotectin assay

Journal

CLINICAL CHEMISTRY AND LABORATORY MEDICINE
Volume 60, Issue 11, Pages 1780-1785

Publisher

WALTER DE GRUYTER GMBH
DOI: 10.1515/cclm-2022-0526

Keywords

calprotectin; inflammatory bowel disease; OC-sensor

Ask authors/readers for more resources

This study evaluated the performance of a new OC-FCa calprotectin assay for the diagnosis of inflammatory bowel disease (IBD) and determined a suitable cutoff value. The results showed that at a cutoff of 600 μg/g, the assay had a sensitivity of 91.8% and specificity of 79.3% for diagnosing IBD.
Objectives Faecal calprotectin (f-Cal) and faecal haemoglobin (f-Hb) are important tests for evaluation of gastrointestinal disease. Samples for measurement of f-Hb are taken by the patient directly into a specimen collection device containing stabilising buffer, which can be placed directly onto the analyser in the laboratory. Samples for f-Cal are usually sent in screw top pots and often require time-consuming extraction procedures prior to analysis. OC-FCa calprotectin is a new assay which uses the same specimen collection device and analyser as our current f-Hb assay. Analytical evaluation has already shown it to perform well but to have a positive bias. This study was a clinical evaluation to investigate the diagnostic test performance and cut-off suitable for its use in the diagnosis of IBD. Methods OC-FCa calprotectin was measured in a convenience sample of 603 patients in whom f-Hb had been requested and was found to be >= 10 mu g/g. Clinical outcomes were obtained from notes, radiological reports and endoscopy and histology reports. Results A total of 425 patients completed clinical investigations; IBD was diagnosed in 49 and other colorectal pathology in 161. Median f-Cal in patients with IBD was 1,660 mu g/g, significantly different (p<0.01) from those with other colorectal pathology (192 mu g/g) or normal findings (157 mu g/g). ROC curve analysis showed AUC of 0.898 with sensitivity of 91.8% and specificity of 79.3% at a cut-off of 600 mu g/g. Conclusions The new OC-FCa calprotectin assay performed well for the diagnosis of IBD using a cut-off of 600 mu g/g.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available