4.8 Article

Flow Field-Flow Fractionation with a Thickness-Tapered Channel

Journal

ANALYTICAL CHEMISTRY
Volume 94, Issue 41, Pages 14460-14466

Publisher

AMER CHEMICAL SOC
DOI: 10.1021/acs.analchem.2c03503

Keywords

-

Funding

  1. National Research Foundation (NRF) of Korea
  2. [NRF-2021R1A2C2003171]

Ask authors/readers for more resources

This study introduces the thickness-tapered channel design for flow field-flow fractionation (FlFFF) which can improve the migration of long-retaining smaller-sized particles and the resolution of high size separation without the need for additional field-programming options.
This study introduces the thickness-tapered channel design for flow field-flow fractionation (FlFFF) for the first time. In this design, the channel thickness linearly decreases along the channel axis such that the flow velocity increases down the channel. Channel thickness is an important variable for controlling retention time and resolution in field-flow fractionation. Especially, in the steric/hyperlayer mode of FlFFF, in which particles (>1 mu m) migrate at elevated heights above the channel wall owing to hydrodynamic lift forces, the migration of long-retaining smaller-sized particles can be enhanced in a relatively thin channel or by increasing the migration flow rate; however, an upper size limit that can be resolved is simultaneously sacrificed. A thickness-tapered channel was constructed without a channel spacer by carving the surface of a channel block such that the channel inlet was deeper than the outlet (w = 400 -> 200 mu m). The performance of a thickness-tapered channel was evaluated using polystyrene standards and compared to that of a channel of uniform thickness (w = 300 mu m) with a similar effective channel volume in terms of sample recovery, dynamic size range of separation, and steric transition under different flow rate conditions. The thickness-tapered channel can be an alternative to maintain the resolving power for particles with an upper large-diameter limit, faster separation of particles with a lower limit, and higher elution recovery without implementing the additional field-programming option.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.8
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available