4.1 Article

Manual versus Mechanical Delivery of High-Quality Cardiopulmonary Resuscitation on a River-Based Fire Rescue Boat

Journal

PREHOSPITAL AND DISASTER MEDICINE
Volume 37, Issue 5, Pages 630-637

Publisher

CAMBRIDGE UNIV PRESS
DOI: 10.1017/S1049023X22001042

Keywords

cardiac arrest; high-quality cardiopulmonary resuscitation; mechanical CPR; prehospital care; river-based CPR

Ask authors/readers for more resources

This study investigates the efficacy of manual and mechanical delivery of high-quality cardiopulmonary resuscitation (HQ-CPR) on a fire rescue boat. The results show that mechanical HQ-CPR is superior to manual HQ-CPR during both stable and dynamic riverine navigation. Whereas manual CPR delivery is worse during dynamic transportation conditions compared to stable conditions, mechanical CPR delivery is unaffected by drive style. This suggests the routine use of mechanical HQ-CPR devices in the riverine patient transport setting.
Objectives: Studies have demonstrated the efficacy of mechanical devices at delivering high-quality cardiopulmonary resuscitation (HQ-CPR) in various transport settings. Herein, this study investigates the efficacy of manual and mechanical HQ-CPR delivery on a fire rescue boat. Methods: A total of 15 active firefighter-paramedics were recruited for a prospective manikin-based trial. Each paramedic performed two minutes manual compression-only CPR while navigating on a river-based fire rescue boat. The boat was piloted in either a stable linear manner or dynamic S-turn manner to simulate obstacle avoidance. For each session of manual HQ-CPR, a session of mechanical HQ-CPR was also performed with a LUCAS 3 (Stryker; Kalamazoo, Michigan USA). A total of 60 sessions were completed. Parameters recorded included compression fraction (CF) and the percentage of compressions with correct depth >5cm (D%), correct rate 100-120 (R%), full release (FR%), and correct hand position (HP%). A composite HQ-CPR score was calculated as follows: ((D% + R% + FR% + HP%)/4) * CF%). Differences in magnitude of change seen in stable versus dynamic navigation within study conditions were evaluated with a Z-score calculation. Difficulty of HQ-CPR delivery was assessed utilizing the Borg Rating of Perceived Exertion Scale. Results: Participants were mostly male and had a median experience of 20 years. Manual HQ-CPR delivered during stable navigation out-performed manual HQ-CPR delivered during dynamic navigation for composite score and trended towards superiority for FR% and R%. There was no difference seen for any measured variable when comparing mechanical HQ-CPR delivered during stable navigation versus dynamic navigation. Mechanical HQ-CPR out-performed manual HQ-CPR during both stable and dynamic navigation in terms of composite score, FR%, and R%. Z-score calculation demonstrated that manual HQ-CPR delivery was significantly more affected by drive style than mechanical HQ-CPR delivery in terms of composite HQ-CPR score and trended towards significance for FR% and R%. Borg Rating of Perceived Exertion was higher for manual CPR delivered during dynamic sessions than for stable sessions. Conclusion: Mechanical HQ-CPR delivery is superior to manual HQ-CPR delivery during both stable and dynamic riverine navigation. Whereas manual HQ-CPR delivery was worse during dynamic transportation conditions compared to stable transport conditions, mechanical HQ-CPR delivery was unaffected by drive style. This suggests the utility of routine use of mechanical HQ-CPR devices in the riverine patient transport setting.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.1
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available