4.7 Article

Simple sacrificial-layer-free microfabrication processes for air-cavity Fresnel acoustic lenses (ACFALs) with improved focusing performance

Journal

MICROSYSTEMS & NANOENGINEERING
Volume 8, Issue 1, Pages -

Publisher

SPRINGERNATURE
DOI: 10.1038/s41378-022-00407-w

Keywords

-

Funding

  1. National Institutes of Health [1R01 EB026284]

Ask authors/readers for more resources

This paper introduces new and improved types of acoustic lenses (ACFALs) based on different materials, manufactured through a simple sacrificial-layer-free microfabrication process. The effect of lens thickness on acoustic transmittance is studied and thickness control techniques are optimized. These techniques are useful for manufacturing high-performance ACFALs and other miniaturized devices.
Focused ultrasound (FUS) is a powerful tool widely used in biomedical therapy and imaging as well as in sensors and actuators. Conventional focusing techniques based on curved surfaces, metamaterial structures, and multielement phased arrays either present difficulties in massively parallel manufacturing with high precision or require complex drive electronics to operate. These difficulties have been addressed by microfabricated self-focusing acoustic transducers (SFATs) with Parylene air-cavity Fresnel acoustic lenses (ACFALs), which require a time-demanding step in removing the sacrificial layer. This paper presents three new and improved types of ACFALs based on polydimethylsiloxane (PDMS), an SU-8/PDMS bilayer, and SU-8, which are manufactured through simple sacrificial-layer-free microfabrication processes that are two to four times faster than that for the Parylene ACFALs. Moreover, by studying the effect of the lens thickness on the acoustic transmittance through the lens, the performance of the transducers has been optimized with improved thickness control techniques developed for PDMS and SU-8. As a result, the measured power transfer efficiency (PTE) and peak output acoustic pressure are up to 2.0 and 1.8 times higher than those of the Parylene ACFALs, respectively. The simple microfabrication techniques described in this paper are useful for manufacturing not only high-performance ACFALs but also other miniaturized devices with hollow or suspended structures for microfluidic and optical applications.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available