4.4 Article

Effect of specimen processing for transmission electron microscopy on lattice spacing variation in Si specimens

Journal

ULTRAMICROSCOPY
Volume 238, Issue -, Pages -

Publisher

ELSEVIER
DOI: 10.1016/j.ultramic.2022.113537

Keywords

Transmission electron microscopy; Metrology; Lattice spacing; Specimen processing; Silicon

Categories

Funding

  1. JSPS KAKENHI [20K04511]
  2. Grants-in-Aid for Scientific Research [20K04511] Funding Source: KAKEN

Ask authors/readers for more resources

The variation in the (220) lattice spacing of Si due to specimen processing for TEM has been experimentally evaluated, showing that the processing conditions and measurement location can significantly affect the measurement uncertainty in sub-nanometer metrology using TEM.
Variation in the (220) lattice spacing of Si due to specimen processing for transmission electron microscopy (TEM) was experimentally evaluated by comparing the measured lattice spacings of crystalline specimens processed by crushing, mechanical polishing only, and combined mechanical and Ar ion polishing. Although distinct variation in the (220) lattice spacing between the Si specimens processed by crushing and by mechanical polishing only is not observed, the (220) lattice spacing of specimens prepared by combined mechanical and Ar ion polishing imply increasing tendency with increasing Ar ion beam irradiation time. Moreover, the (220) lattice spacing measured from the outermost region of the specimen edge tends to be approximately 3% to 5% larger than that measured from the inner region, irrespective of the processing method. These results demonstrate that differences in the processing conditions of Si specimens and in the measurement location of the Si lattice spacing can be major component of the measurement uncertainty in sub-nanometer metrology using TEM with magnification calibration by the Si lattice spacing. When attempting to apply the lattice spacing of Si as a reference with traceability to the International System of Units for TEM magnification calibration in sub-nanometer metrology, the results suggest that the effect of specimen processing on variation in the lattice spacing is not negligible.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.4
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available