Challenges in measuring nitrogen isotope signatures in inorganic nitrogen forms: An interlaboratory comparison of three common measurement approaches

Journal

RAPID COMMUNICATIONS IN MASS SPECTROMETRY
Volume 36, Issue 22, Pages -

Publisher

WILEY
DOI: 10.1002/rcm.9370

Keywords

-

Funding

  1. European Association of National Metrology Institutes [16ENV06 SIRS, 19ENV05 STELLAR]

Ask authors/readers for more resources

This study conducted an interlaboratory comparison of different methods for measuring delta N-15 in NO3- and NH4+ in 10 European laboratories. The results showed that laboratories using CM-N2O performed superior for both NO3- and NH4+, followed by DN. Most methods and laboratories underestimated the at%N-15 of N-i of labeled standards, but relative errors were within maximal 6% deviation from the real value and therefore acceptable.
Rationale Stable isotope approaches are increasingly applied to better understand the cycling of inorganic nitrogen (N-i) forms, key limiting nutrients in terrestrial and aquatic ecosystems. A systematic comparison of the accuracy and precision of the most commonly used methods to analyze delta N-15 in NO3- and NH4+ and interlaboratory comparison tests to evaluate the comparability of isotope results between laboratories are, however, still lacking. Methods Here, we conducted an interlaboratory comparison involving 10 European laboratories to compare different methods and laboratory performance to measure delta N-15 in NO3- and NH4+. The approaches tested were (a) microdiffusion (MD), (b) chemical conversion (CM), which transforms N-i to either N2O (CM-N2O) or N-2 (CM-N-2), and (c) the denitrifier (DN) methods. Results The study showed that standards in their single forms were reasonably replicated by the different methods and laboratories, with laboratories applying CM-N2O performing superior for both NO3- and NH4+, followed by DN. Laboratories using MD significantly underestimated the true values due to incomplete recovery and also those using CM-N-2 showed issues with isotope fractionation. Most methods and laboratories underestimated the at%N-15 of N-i of labeled standards in their single forms, but relative errors were within maximal 6% deviation from the real value and therefore acceptable. The results showed further that MD is strongly biased by nonspecificity. The results of the environmental samples were generally highly variable, with standard deviations (SD) of up to +/- 8.4 parts per thousand for NO3- and +/- 32.9 parts per thousand for NH4+; SDs within laboratories were found to be considerably lower (on average 3.1 parts per thousand). The variability could not be connected to any single factor but next to errors due to blank contamination, isotope normalization, and fractionation, and also matrix effects and analytical errors have to be considered. Conclusions The inconsistency among all methods and laboratories raises concern about reported delta N-15 values particularly from environmental samples.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.4
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available