A novel multichannel deep learning model for fast denoising of Monte Carlo dose calculations: preclinical applications

Journal

PHYSICS IN MEDICINE AND BIOLOGY
Volume 67, Issue 16, Pages -

Publisher

IOP Publishing Ltd
DOI: 10.1088/1361-6560/ac8390

Keywords

Monte Carlo dose calculation; preclinical radiotherapy; Monte Carlo uncertainty; denoising; deep learning

Ask authors/readers for more resources

This study proposes a novel deep learning pipeline for denoising preclinical kV dose distributions. The model achieves high precision dose distributions by using low precision Monte Carlo dose calculations, Monte Carlo uncertainty and mass density maps as additional inputs. The proposed model reduces computation time significantly and shows promising results in accuracy.
Objective. In preclinical radiotherapy with kilovolt (kV) x-ray beams, accurate treatment planning is needed to improve the translation potential to clinical trials. Monte Carlo based radiation transport simulations are the gold standard to calculate the absorbed dose distribution in external beam radiotherapy. However, these simulations are notorious for their long computation time, causing a bottleneck in the workflow. Previous studies have used deep learning models to speed up these simulations for clinical megavolt (MV) beams. For kV beams, dose distributions are more affected by tissue type than for MV beams, leading to steep dose gradients. This study aims to speed up preclinical kV dose simulations by proposing a novel deep learning pipeline. Approach. A deep learning model is proposed that denoises low precision (similar to 10(6) simulated particles) dose distributions to produce high precision (10(9) simulated particles) dose distributions. To effectively denoise the steep dose gradients in preclinical kV dose distributions, the model uses the novel approach to use the low precision Monte Carlo dose calculation as well as the Monte Carlo uncertainty (MCU) map and the mass density map as additional input channels. The model was trained on a large synthetic dataset and tested on a real dataset with a different data distribution. To keep model inference time to a minimum, a novel method for inference optimization was developed as well. Main results. The proposed model provides dose distributions which achieve a median gamma pass rate (3%/0.3 mm) of 98% with a lower bound of 95% when compared to the high precision Monte Carlo dose distributions from the test set, which represents a different dataset distribution than the training set. Using the proposed model together with the novel inference optimization method, the total computation time was reduced from approximately 45 min to less than six seconds on average. Significance. This study presents the first model that can denoise preclinical kV instead of clinical MV Monte Carlo dose distributions. This was achieved by using the MCU and mass density maps as additional model inputs. Additionally, this study shows that training such a model on a synthetic dataset is not only a viable option, but even increases the generalization of the model compared to training on real data due to the sheer size and variety of the synthetic dataset. The application of this model will enable speeding up treatment plan optimization in the preclinical workflow.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available