4.6 Article

Developing breast lesion detection algorithms for digital breast tomosynthesis: Leveraging false positive findings

Journal

MEDICAL PHYSICS
Volume 49, Issue 12, Pages 7596-7608

Publisher

WILEY
DOI: 10.1002/mp.15883

Keywords

breast cancer; computer-aided detection; deep learning; digital breast tomosynthesis; lesion detection

Funding

  1. National Institutes of Health [R37 CA248207]

Ask authors/readers for more resources

This study developed algorithms for detecting biopsy-proven breast lesions on digital breast tomosynthesis (DBT) using multi-depth level convolutional models and leveraging non-biopsied samples. The study found that false positive findings from non-biopsied benign lesions could improve detection algorithms, and the ensemble model with multi-depth levels showed improved lesion detection performance.
Background Due to the complex nature of digital breast tomosynthesis (DBT) in imaging techniques, reading times are longer than 2D mammograms. A robust computer-aided diagnosis system in DBT could help radiologists reduce their workload and reading times. Purpose The purpose of this study was to develop algorithms for detecting biopsy-proven breast lesions on DBT using multi-depth level convolutional models and leveraging non-biopsied samples. As biopsied positive samples in a lesion dataset are limited, we hypothesized that false positive (FP) findings by detection algorithms from non-biopsied benign lesions could improve detection algorithms by using them as data augmentation. Approach We first extracted 2D slices from DBT volumes with biopsy-proven breast lesions (cancer and benign), with non-biopsied benign lesions (actionable), and for controls. Then, to provide lesion continuity along the z-direction, we combined a lesion slice with its immediate adjacent slices to synthesize 2.5-dimensional (2.5D) images of the lesion by assigning them into R, G, and B color channels. We used 224 biopsy-proven lesions from 39 cancer and 62 benign patients from a DBTex challenge dataset of 1000 scans. We included the 2.5D images of immediate neighboring slices from the lesion's center to increase the number of training samples. For lesion detection, we used the YOLOv5 algorithm as our base network. We trained a baseline algorithm (medium-depth level) using biopsied samples to detect actionable FPs in non-biopsied images. Afterward, we fine-tuned the baseline model on the augmented image set (actionable FPs added). For lesion inferencing, we processed the DBT volume slice-by-slice to estimate bounding boxes in each slice, and then combined them by connecting bounding boxes along the depth via volumetric morphological closing. We trained an additional model (large) with deeper-depth levels by repeating the above process. Finally, we developed an ensemble algorithm by combining the medium and large detection models. We used the free-response operating characteristic curve to evaluate our algorithms. We reported mean sensitivity per FPs per DBT volume only for biopsied views and sensitivity at 2-false positives per image (2FPI) for all views. However, due to the limited accessibility to the truth of the challenge validation and test datasets, we used sensitivity at 2FPI for statistical evaluation. Results For the DBTex independent validation set, the medium baseline model achieved a mean sensitivity of 0.627 FPs per DBT volume, and a sensitivity of 0.640 at 2FPI. After adding actionable FP lesions, the model had an improved 2FPI of 0.769 over the baseline (p-value = 0.013). Our ensemble algorithm with multi-depth levels (medium + large) achieved a mean sensitivity of 0.815 FPs per DBT volume and an improved sensitivity at 2FPI of 0.80 over the baseline (p-value < 0.001) on the validation set. Finally, our ensemble model achieved a mean sensitivity of 0.786 FPs per DBT volume and a sensitivity of 0.743 at 2FPI on the DBTex independent test set. Conclusions Our results show that actionable FP findings hold useful information for lesion detection algorithms, and our ensemble detection model with multi-depth levels improves lesion detection performance.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available