4.6 Article

Comparison of four different immunoassays and a rapid isotope-dilution liquid chromatography-tandem mass spectrometry assay for serum folate

Journal

CLINICAL CHEMISTRY AND LABORATORY MEDICINE
Volume 60, Issue 9, Pages 1393-1402

Publisher

WALTER DE GRUYTER GMBH
DOI: 10.1515/cclm-2021-1283

Keywords

immunoassays; isotope-dilution liquid chromatography-tandem mass spectrometry (ID-LC-MS; MS); method comparison; serum folate

Funding

  1. National Key Research and Development Program of China [2018YFC1002204]

Ask authors/readers for more resources

This study compared the differences between different methods for measuring serum folate and found significant variations among the assays. Although there was good correlation between the assays, there was a lack of agreement. Therefore, it is important to routinely assess the accuracy and consistency of serum folate assays and promote standardization efforts.
Objectives Accurate measurement of serum folate is essential for the diagnosis and management of various disorders. This study aims to investigate the between-method differences of four immunoassays and a rapid isotope-dilution liquid chromatography-tandem mass spectrometry (ID-LC-MS/MS) method. Methods Roche Cobas (USA), Abbott Alinity i2000 (USA), Beckman Coulter Access (USA), Mindray CL-6000i (China), and the ID-LC-MS/MS method were compared using 46 human serum samples. The results were analysed by Passing-Bablok regressions and Bland-Altman plots. A bias of 13.31% based on biological variation was used as the bias criterion. Results All the within-run and total coefficients of variation (CVs) met the specification. The folate concentrations determined by all the assays were significantly different (p=0.0028). All assays had correlation coefficients over 0.97 with each other. The 95% confidence intervals (CIs) for the slope seldom contained 1 and few 95% CIs for the intercept contained 0 in the regression equations. Compared to ID-LC-MS/MS, the biases of all assays ranged from -20.91 to 13.56 nmol/L, and the mean relative biases ranged from -9.85 to 40.33%. The predicted mean relative biases at the medical decision levels rarely met the criterion. Conclusions Assays for serum folate had good correlations with each other but lacked good agreement. The accuracy and consistency of assays for serum folate should be measured and assessed routinely. Standardization work to improve the accuracy of serum folate assays, such as the extension of traceability to reference methods or materials, calibration standardization efforts, and assay-adjusted cut-offs should be promoted.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available