4.5 Article

Storage conditions, sample integrity, interferences, and a decision tool for investigating unusual high-sensitivity cardiac troponin results

Journal

CLINICAL BIOCHEMISTRY
Volume 115, Issue -, Pages 67-76

Publisher

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.clinbiochem.2022.06.007

Keywords

High-sensitivity cardiac troponin; Stability; Matrix; Interference; Accuracy; Pre-analytical

Ask authors/readers for more resources

The current definition and usage of high-sensitivity cardiac troponin (hs-cTn) assays have drawn significant attention in the literature due to the different algorithms and pathways used for patient care. This literature review focuses on the impact of pre-analytical conditions and interferences on hs-cTn concentrations and provides a breakdown of studies conducted by major diagnostic companies. It highlights the importance of pre-analytical variables in interpreting hs-cTn assays and suggests the need for further studies in this area.
The current definition of high-sensitivity cardiac troponin (hs-cTn) assays is laboratory-based and their analytical attributes and characteristics have drawn significant attention in the literature at least partly due to the lower concentration cut-offs and changes in concentrations (i.e., deltas) employed in different algorithms and pathways to manage patient care. We propose that pre-analytical conditions such as sample type, storage conditions, and other interferences may also have a significant impact on hs-cTn concentrations and clinical management. The purpose of this literature review is to provide a summary of important pre-analytical and interference studies affecting hs-cTn concentrations. A breakdown of the literature for the major diagnostic companies providing core laboratory instrumentation (i.e., Abbott, Beckman, Ortho, Roche, and Siemens) is also provided. Finally, three cases are highlighted where knowledge of pre-analytical factors aids the hs-cTn clinically discordant investigations. This review highlights the importance of pre-analytical variables, especially storage condition, sample handling, and blood tubes used (i.e., sample type) when interpreting hs-cTn assays. Additional studies are needed to further elaborate on pre-analytical variables (i.e., centrifugation, sample type, stability) and interferences for all hs-cTn assays in clinical use, as knowledge of these variables may aid in hs-cTn clinically discordant investigations.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available