4.7 Article

Direct Comparison of Alere i and cobas Liat Influenza A and B Tests for Rapid Detection of Influenza Virus Infection

Journal

JOURNAL OF CLINICAL MICROBIOLOGY
Volume 54, Issue 11, Pages 2763-2766

Publisher

AMER SOC MICROBIOLOGY
DOI: 10.1128/JCM.01586-16

Keywords

-

Categories

Funding

  1. BioFire Diagnostics

Ask authors/readers for more resources

We compared two rapid, point-of care nucleic acid amplification tests for detection of influenza A and B viruses (Alere i [Alere] and cobas Liat [Roche Diagnostics]) with the influenza A and B virus test components of the FilmArray respiratory panel (Bio-Fire Diagnostics) using 129 respiratory specimens collected in universal viral transport medium (80 influenza A virus and 16 influenza B virus positive) from both adult and pediatric patients. The sensitivities of the Alere test were 71.3% for influenza A virus and 93.3% for influenza B virus, with specificities of 100% for both viruses. The sensitivities and specificities of the Liat test were 100% for both influenza A and B viruses. The poor sensitivity of the Alere test for detection of influenza A virus was likely due to a study set that included many low-positive samples that were below its limit of detection.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available