Visual Reassessment with Flux-Interval Plot Configuration after Automatic Classification for Accurate Atrial Fibrillation Detection by Photoplethysmography

Journal

DIAGNOSTICS
Volume 12, Issue 6, Pages -

Publisher

MDPI
DOI: 10.3390/diagnostics12061304

Keywords

atrial fibrillation; arrhythmia; photoplethysmography; blood flux; reassessment; configuration

Funding

  1. Research Center for Applied Sciences of Academia Sinica, Taiwan [3010]

Ask authors/readers for more resources

This study developed an automated detection method for atrial fibrillation (AFib) based on photoplethysmogram (PPG) signals and visualized the results for reassessment. The method showed high sensitivity, specificity, accuracy, and precision across 460 samples. The findings demonstrate the potential of this user-friendly technology for in-house AFib diagnostics.
Atrial fibrillation (AFib) is a common type of arrhythmia that is often clinically asymptomatic, which increases the risk of stroke significantly but can be prevented with anticoagulation. The photoplethysmogram (PPG) has recently attracted a lot of attention as a surrogate for electrocardiography (ECG) on atrial fibrillation (AFib) detection, with its out-of-hospital usability for rapid screening or long-term monitoring. Previous studies on AFib detection via PPG signals have achieved good results, but were short of intuitive criteria like ECG p-wave absence or not, especially while using interval randomness to detect AFib suffering from conjunction with premature contractions (PAC/PVC). In this study, we newly developed a PPG flux (pulse amplitude) and interval plots-based methodology, simply comprising an irregularity index threshold of 20 and regression error threshold of 0.06 for the precise automatic detection of AFib. The proposed method with automated detection on AFib shows a combined sensitivity, specificity, accuracy, and precision of 1, 0.995, 0.995, and 0.952 across the 460 samples. Furthermore, the flux-interval plot configuration also acts as a very intuitive tool for visual reassessment to confirm the automatic detection of AFib by its distinctive plot pattern compared to other cardiac rhythms. The study demonstrated that exclusive 2 false-positive cases could be corrected after the reassessment. With the methodology's background theory well established, the detection process automated and visualized, and the PPG sensors already extensively used, this technology is very user-friendly and convincing for promoted to in-house AFib diagnostics.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available