4.1 Article

Mechanical Analysis of 3 Posterior Fusion Assemblies Intended to Cross the Cervicothoracic Junction

Journal

CLINICAL SPINE SURGERY
Volume 35, Issue 4, Pages 144-148

Publisher

LIPPINCOTT WILLIAMS & WILKINS
DOI: 10.1097/BSD.0000000000001317

Keywords

cervicothoracic fusion; cervical spine; spinal fusion; ASTM F1717; thoracic spine; posterior spinal fusion implant

Funding

  1. UAMS Barton Pilot Grant
  2. Medtronic Sofamor Danek USA Inc., Memphis, TN

Ask authors/readers for more resources

This study evaluated the mechanical properties of three posterior fusion assemblies commonly used to cross the cervicothoracic junction. The results showed that the assemblies with a 5.5 mm rod were stiffer compared to those with only a 3.5 mm rod.
Study Design: This was a biomechanical comparison study. Objective: The objective of this study is to evaluate the mechanical properties of 3 posterior spinal fusion assemblies commonly used to cross the cervicothoracic junction. Summary of Background: When posterior cervical fusions are extended into the thoracic spine, an instrumentation transition is often utilized. The cervical rod (3.5 mm) can continue using thoracic screws designed to accept the cervical rods. Alternatively, traditional thoracic screws may be used to accept thoracic rods (5.5 mm). This requires the use of a 3.5-5.5 mm transition rod or a separate 5.5 mm rod and a connector to fix the 3.5 and 5.5 mm rod together. Fusion success depends on the immobilization of vertebrae, yet the mechanics provided by these different assemblies are unknown. Materials and Methods: Three titanium alloy posterior fusion assemblies intended to cross the cervicothoracic junction underwent static compressive bending, tensile bending, and torsion as described in ASTM F1717 to a torque of 2.5 Nm. Five samples of each assembly were attached to ultrahigh molecular weight polyethylene blocks via multiaxial screws for testing. Force and displacement were recorded, and the stiffness of each construct was calculated. Results: The 2 assemblies that included a 5.5 mm rod were found to be stiffer and have less range of motion than the assembly that used only 3.5 mm rods. Conclusions: The results of this study indicate that incorporating a 5.5 mm rod in a fusion assembly adds significant stiffness to the construct. When the stability of a fusion is of heightened concern, as demonstrated by the ASTM F1717 vertebrectomy (worst-case scenario) model, including 5.5 mm rods may increase fusion success rates.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.1
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available