Rapid Antigen Test LumiraDxTM vs. Real Time Polymerase Chain Reaction for the Diagnosis of SARS-CoV-2 Infection: A Retrospective Cohort Study

Publisher

MDPI
DOI: 10.3390/ijerph19073826

Keywords

SARS-CoV-2; COVID-19; diagnosis; rapid antigen testing; RT-PCR; agreement; sensitivity; specificity; positive predictive value; negative predictive value; accuracy

Ask authors/readers for more resources

This study suggests that the LumiraDx(TM) SARS-CoV-2 antigen assay may be suitable for detecting SARS-CoV-2 infection, especially in the early phase. Compared to real time RT-PCR, LumiraDx(TM) has lower sensitivity, but sensitivity increases when testing is conducted within 10 days of symptom onset and with Ct values below 25.
Background: Real time reverse transcription polymerase chain reaction (real time RT-PCR) testing is the gold standard for the diagnosis of SARS-CoV-2 infections. However, to expand the testing capacity, new SARS-CoV-2 rapid antigen tests (Ag-RDTs) have been implemented. Ag-RDTs are more rapid, but less reliable in terms of sensitivity, and real-life data on their performance in comparison with the real time RT-PCR test are lacking. Methods: We aimed at assessing the diagnostic performance of the third-generation antigenic swab LumiraDx (TM) compared with real time RT-PCR in a retrospective cohort study at the Infectious Diseases Unit of Padua. All of the patients who were consecutively tested for SARS-CoV-2 in our centre (by both real time RT-PCR and Ag-RTD LumiraDx(TM)) from 19 January to 30 May 2021, were included. Cycle-threshold (Ct) values of positive real time RT-PCR were recorded as well as the number of days from symptoms' onset to testing. Results: Among the 282 patients included, 80.9% (N = 228) tested positive to real time RT-PCR, and among these, 174 tested positive also to LumiraDx (TM). Compared with real time RT-PCR, which is considered as the gold standard for the assessment of the presence/absence of SARS-CoV-2 infection, LumiraDx (TM) showed an overall sensitivity of 76.3% and specificity of 94.4%. Sensitivity increased to 91% when testing was performed <10 days from symptoms' onset, and to 95% when considering Ct < 25. Multivariable binomial logistic regression showed that false negative LumiraDx (TM) results were significantly associated with high Ct values, and with further testing from symptoms' onset. Conclusions: The results of our study suggested that the LumiraDx (TM) SARS-CoV-2 antigen assay may be appropriate for the detection of SARS-CoV-2 infection, especially in its early phase when the test largely meets the performance requirements of the European Centre for Disease Prevention and Control (ECDC).

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.3
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available