Cardiac radioablation for ventricular tachycardia: Which approach for incorporating cardiorespiratory motions into the planning target volume

Journal

Publisher

ELSEVIER SCI LTD
DOI: 10.1016/j.ejmp.2022.01.004

Keywords

Cardiac radioablation; Ventricular tachycardia; Motion management; ITV; 4D CT

Funding

  1. French National Research Association (ANR) [ANR-11-LABX-0004]
  2. French Federation of Cardiology (FFC)

Ask authors/readers for more resources

The study evaluates different approaches for generating a cardiorespiratory internal target volume (ITV) for cardiac radioablation. The results suggest that combining two extreme cardiac phases with 10 respiratory phases is a robust approach, while using a single cardiac phase is not recommended.
Purpose: To evaluate different approaches for generating a cardiorespiratory ITV for cardiac radioablation. Methods: Four patients with ventricular tachycardia were included in this study. For each patient, cardiac-gated and respiration-correlated 4D-CT scans were acquired. The cardiorespiratory ITV was defined using registrations of the cardiac and respiratory 4D-CT images. Five different approaches, which differed in the number of incorporated cardiac phases (1, 2, 10, or 1 with a fixed 3 mm margin (FM) expansion) and respiratory phases (2 or 10), were evaluated. For each approach, a VMAT treatment plan was simulated. Target coverage (TC) and spill were evaluated geometrically and dosimetrically for each approach. Results: When employing one cardiac phase, the TC did not exceed 85%. Using the two extreme phases of the cardiac and respiratory cycles resulted in a geometric TC < 88% for two patients, with a dosimetric TC of 83% for one patient. An acceptable TC for all patients (geometric TC > 89%, dosimetric TC > 92%) was only achieved when combining 10 respiratory phases with either 2 or 10 cardiac phases or a single cardiac phase with FM. The use of a single cardiac phase with FM combined with 10 respiratory phases lead to a mean geometric and dosimetric spill of 43% and 35%, respectively. Conclusion: For cardiac radioablation, the use of two extreme cardiac phases combined with 10 respiratory phases is a robust approach to generate a cardiorespiratory ITV. The use of a single cardiac phase with or without fixed margin expansion is not recommended based on this study.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.4
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available