Ask authors/readers for more resources

This study investigated the use of Z-stacked images for diagnostic assessment of urine ThinPrep WSI. The results show that Z-stack images provide minimal diagnostic benefit for urine ThinPrep WSI and do not justify the prolonged scan times and larger storage needs compared to those without Z-stack.
BACKGROUND Whole slide imaging (WSI) adoption has been slower in cytopathology due, in part, to challenges in multifocal plane scanning on 3-dimensional cell clusters. ThinPrep and other liquid-based preparations may alleviate the issue by reducing clusters in a concentrated area. This study investigates the use of Z-stacked images for diagnostic assessment and the experience of evaluating urine ThinPrep WSI. METHODS Thirty ThinPrep urine cases of high-grade urothelial carcinoma (n = 22) and cases of negative for high-grade urothelial carcinoma (n = 8) were included. Slides were scanned at 40x magnification without Z-stack and with Z-stack at 3 layers, 1 mu m each. Six cytopathologists and 1 cytotechnologist evaluated the cases in 2 rounds with a 2-week wash-out period in a blinded manner. A Cohen's Kappa (CK) calculated concordance rates. A survey after each round evaluated participant experience. RESULTS CK with the original report ranged from 0.606 to 1.0 (P < .05) without Z-stack and 0.533 to 1.0 (P < .05) with Z-stack both indicating substantial-to-perfect concordance. For both rounds, interobserver CK was moderate-to-perfect (0.417-1.0, P < .05). Intraobserver CK was 0.697-1.0 (P < 0.05), indicating substantial to perfect concordance. The average scan time and file size for slides without Z-stack and with Z-stack are 6.27 minute/0.827 GB and 14.06 minute/2.650 GB, respectively. Surveys demonstrated a range in comfort and use with slightly more favorable opinions for Z-stacked cases. CONCLUSIONS Z-stack images provide minimal diagnostic benefit for urine ThinPrep WSI. In addition, Z-stacked urine WSI does not justify the prolonged scan times and larger storage needs compared to those without Z-stack.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.4
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available