Performance of Wide-Angle Tomosynthesis with Synthetic Mammography in Comparison to Full Field Digital Mammography

Journal

ACADEMIC RADIOLOGY
Volume 30, Issue 1, Pages 3-13

Publisher

ELSEVIER SCIENCE INC
DOI: 10.1016/j.acra.2022.03.026

Keywords

Tomosynthesis; wide-angle; synthetic mammography; full-field digital mammography

Ask authors/readers for more resources

This study aimed to compare the superiority of wide-angle digital breast tomosynthesis plus synthetic mammography (Insight 2D) to full-field digital mammography (FFDM). The results showed that wide-angle DBT plus Insight 2D outperformed FFDM in terms of higher accuracy, sensitivity, and lower recall rates in diagnosing breast cancer.
Rationale and Objectives: The purpose of this study was to test for superiority of wide-angle digital breast tomosynthesis plus synthetic mammography (Insight 2D) in comparison to full-field digital mammography (FFDM).Materials and Methods: In this study, twenty readers interpreted 350 screening and diagnostic cases of wide-angle digital breast tomosynthesis (DBT) plus Insight 2D and FFDM in two separate reading sessions separated by at least a 6-week washout period. Breast-level estimates of the area under the curve and sensitivity along with subject-level recall rate were measured and compared between wideangle DBT plus Insight 2D and FFDM. The same measures were also assessed for dense breasts. A hierarchical analysis plan was used to control the study's type I error rate at 0.05. Results: The mean breast-level area under the curve for distinguishing breasts with cancer from non-cancer breasts was 0.893 with DBT plus Insight 2D versus 0.837 with FFDM, showing superiority of DBT plus Insight 2D (p < 0.001). Breast-level sensitivity was significantly superior for DBT plus Insight 2D in comparison to FFDM (0.852 vs. 0.805, p = 0.043). Subject-level recall rate for DBT plus Insight 2D was significantly lower in comparison to FFDM (0.344 vs. 0.473, p < 0.001). For dense breasts, the readers' accuracy with DBT plus Insight 2D was superior to their accuracy with FFDM (0.875 vs. 0.830, p = 0.026), and their recall rate was significantly lower for DBT plus Insight 2D in comparison to FFDM (0.338 vs. 0.441, p = 0.003).Conclusion: Reader performance with wide-angle DBT plus Insight 2D is superior to that with FFDM, showing significantly higher breastlevel accuracy and sensitivity and significantly lower recall rates.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available