Critical Drug Loss Induced by Silicone and Polyurethane Implantable Catheters in a Simulated Infusion Setup with Three Model Drugs

Journal

PHARMACEUTICS
Volume 13, Issue 10, Pages -

Publisher

MDPI
DOI: 10.3390/pharmaceutics13101709

Keywords

infusion line; drug loss; drug analysis; sorption; catheters; surface characterization

Ask authors/readers for more resources

The study experimentally investigated the drug loss caused by silicone and polyurethane implantable catheters, finding that polyurethane catheters caused the highest losses of diazepam and insulin. Drug concentrations were monitored by liquid chromatography, with no loss of paracetamol observed in the complete setups. Catheters were shown to be the cause of some extremely high drug losses that could not be countered by optimizing the extension set in the setup.
Silicone and polyurethane are biocompatible materials used for the manufacture of implantable catheters, but are known to induce drug loss by sorption, causing potentially important clinical consequences. Despite this, their impact on the drugs infused through them is rarely studied, or they are studied individually and not part of a complete infusion setup. The aim of this work was to experimentally investigate the drug loss that these devices can cause, on their own and within a complete infusion setup. Paracetamol, diazepam, and insulin were chosen as models to assess drug sorption. Four commonly used silicone and polyurethane catheters were studied independently and as part of two different setups composed of a syringe, an extension set, and silicone or polyurethane implantable catheter. Simulated infusion through the catheter alone or through the complete setup were tested, at flowrates of 1 mL/h and 10 mL/h. Drug concentrations were monitored by liquid chromatography, and the silicone and polyurethane materials were characterized by ATR-IR spectroscopy and Zeta surface potential measurements. The losses observed with the complete setups followed the same trend as the losses induced individually by the most sorptive device of the setup. With the complete setups, no loss of paracetamol was observed, but diazepam and insulin maximum losses were respectively of 96.4 & PLUSMN; 0.9% and 54.0 & PLUSMN; 5.6%, when using a polyurethane catheter. Overall, catheters were shown to be the cause of some extremely high drug losses that could not be countered by optimizing the extension set in the setup.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available