A Novel Proton Pencil Beam Scanning FLASH RT Delivery Method Enables Optimal OAR Sparing and Ultra-High Dose Rate Delivery: A Comprehensive Dosimetry Study for Lung Tumors

Journal

CANCERS
Volume 13, Issue 22, Pages -

Publisher

MDPI
DOI: 10.3390/cancers13225790

Keywords

proton therapy; pencil beam scanning; transmission beam; proton Bragg peak FLASH; ultra-high dose rate; FLASH radiotherapy; lung hypofractionation

Categories

Ask authors/readers for more resources

This study compares the performance of single-energy Bragg peak proton beams with transmission proton beams in lung tumor treatment planning, finding that Bragg peak plans achieve superior dosimetry performances in OARs while maintaining comparable dose rate performances. Beam angle optimization further improves OAR dosimetry parameters without compromising the 3D FLASH dose rate coverage.
Simple SummaryThis study attempts to answer a novel and clinically relevant question of the value of Bragg-peak-based FLASH planning for lung tumors. Most existing studies and literature are limited to using transmission proton beams at ultra-high dose rates, resulting in unnecessary irradiation exposure to normal tissues beyond the target volume. By combining a new hardware design (universal range shifter and range compensator) and an inverse planning system, the novel Bragg peak method makes the Bragg-peak-based FLASH planning possible. The treatment planning study and dosimetry comparison between single-energy proton Bragg peak beams and transmission proton beams demonstrated superior performances in OAR sparing and comparable FLASH dose rate of the Bragg peak FLASH. Beam angle optimization can further improve Bragg peak FLASH dosimetry performance while maintaining the similar 3D FLASH dose rate coverage for OARs.Purpose: While transmission proton beams have been demonstrated to achieve ultra-high dose rate FLASH therapy delivery, they are unable to spare normal tissues distal to the target. This study aims to compare FLASH treatment planning using single energy Bragg peak proton beams versus transmission proton beams in lung tumors and to evaluate Bragg peak plan optimization, characterize plan quality, and quantify organ-at-risk (OAR) sparing. Materials and Methods: Both Bragg peak and transmission plans were optimized using an in-house platform for 10 consecutive lung patients previously treated with proton stereotactic body radiation therapy (SBRT). To bring the dose rate up to the FLASH-RT threshold, Bragg peak plans with a minimum MU/spot of 1200 and transmission plans with a minimum MU/spot of 400 were developed. Two common prescriptions, 34 Gy in 1 fraction and 54 Gy in 3 fractions, were studied with the same beam arrangement for both Bragg peak and transmission plans (n = 40 plans). RTOG 0915 dosimetry metrics and dose rate metrics based on different dose rate calculations, including average dose rate (ADR), dose-averaged dose rate (DADR), and dose threshold dose rate (DTDR), were investigated. We then evaluated the effect of beam angular optimization on the Bragg peak plans to explore the potential for superior OAR sparing. Results: Bragg peak plans significantly reduced doses to several OAR dose parameters, including lung V-7.4Gy and V-7Gy by 32.0% (p < 0.01) and 30.4% (p < 0.01) for 34Gy/fx plans, respectively; and by 40.8% (p < 0.01) and 41.2% (p < 0.01) for 18Gy/fx plans, respectively, compared with transmission plans. Bragg peak plans have ~3% less in DADR and ~10% differences in mean OARs in DTDR and DADR relative to transmission plans due to the larger portion of lower dose regions of Bragg peak plans. With angular optimization, optimized Bragg peak plans can further reduce the lung V-7Gy by 20.7% (p < 0.01) and V-7.4Gy by 19.7% (p < 0.01) compared with Bragg peak plans without angular optimization while achieving a similar 3D dose rate distribution. Conclusion: The single-energy Bragg peak plans achieve superior dosimetry performances in OARs to transmission plans with comparable dose rate performances for lung cancer FLASH therapy. Beam angle optimization can further improve the OAR dosimetry parameters with similar 3D FLASH dose rate coverage.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available