New semiconductor detectors based on CdZnTe and CeBr3 crystals were tested for their energy resolution and detection efficiency. A spectrometer facility with these detectors, along with Si for alpha-gamma measurements, was assembled and verified for testing. Results showed that the detectors can be used for studying nuclear reactions and properties of reaction products.
New semiconductor detectors based on CdZnTe and CeBr3 crystals are tested. A spectrometer facility consisting of CdZnTe, CeBr3, and Si (for alpha-gamma measurements) detectors is assembled for verification tests. The main parameters of the detectors such as energy resolution FWHM/E-gamma [keV] and detection efficiency (%) are investigated and measured. The possibilities of measuring photon radiation in a wide energy range are verified. gamma-gamma and alpha-gamma coincidences are measured using the detectors. The results of the measurements show that the detectors can be used to study nuclear reactions and properties of reaction products.