A LAMP-based system for rapid detection of eight common pathogens causing lower respiratory tract infections

Journal

JOURNAL OF MICROBIOLOGICAL METHODS
Volume 190, Issue -, Pages -

Publisher

ELSEVIER
DOI: 10.1016/j.mimet.2021.106339

Keywords

LAMP; Standard culture method; Sanger sequencing; High-throughput sequencing; Pathogenic bacteria; Lower respiratory tract infections

Ask authors/readers for more resources

The study developed an extraction-free LAMP method for rapid detection of eight lower respiratory pathogens, showing high consistency with sequencing methods. Through analysis of 528 samples, the LAMP assay demonstrated higher sensitivity than standard culture methods, with high specificity and sensitivity in diagnosing LRTIs.
Lower respiratory tract infections (LRTIs) are a leading cause of morbidity and mortality worldwide and lack a rapid diagnostic method. To improve the diagnosis of LRTIs, we established an available loop-mediated isothermal amplification (LAMP) assay for the detection of eight common lower respiratory pathogens, including Klebsiella pneumoniae, Pseudomonas aeruginosa, Acinetobacter baumannii, Staphylococcus aureus, Escherichia coli, Haemophilus influenzae, Streptococcus pneumoniae, and Moraxella catarrhalis. The whole process can be achieved within 1 h (sample to results read out). We established an extraction free isothermal system. 528 sputum samples collected from patients suspected to have LRTIs were analyzed by the system (8 tests in each sample, a total of 4224 tests) and compared with the standard culture method (SCM). The samples with inconsistent results were further verified by Sanger sequencing and High-throughput sequencing (NGS). The detection limits of the LAMP assay for the 8 pathogens ranged from 103 to 104 CFU/mL. Upon testing 528 samples, the Kappa coefficients of all pathogens ranged between 0.5 and 0.7 indicated a moderate agreement between the LAMP assay and the SCM. All inconsistent samples were further verified by Sanger sequencing, we found that the developed LAMP assay had a higher consistency level with Sanger sequencing than the SCM for all pathogens. Additionally, when the NGS was set to a diagnostic gold standard, the specificity and sensitivity of the LAMP assay for LRTIs were 94.49% and 75.00%. The present study demonstrated that the developed LAMP has high consistency with the sequencing methods. Meanwhile, the LAMP assay has a higher detection rate compared to the SCM. It may be a powerful tool for rapid and reliable clinical diagnosis of LRTIs in primary hospitals.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.3
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available