Insufficient sampling frequencies skew heart rate variability estimates: Implications for extracting heart rate metrics from neuroimaging and physiological data

Journal

JOURNAL OF BIOMEDICAL INFORMATICS
Volume 123, Issue -, Pages -

Publisher

ACADEMIC PRESS INC ELSEVIER SCIENCE
DOI: 10.1016/j.jbi.2021.103934

Keywords

Heart rate variability; Sampling frequency; Electrocardiography; Validity; Downsampling

Funding

  1. Canadian Institutes of Health Research [173416]
  2. University of Calgary
  3. Natural Sciences and Engineering Research Council [CGSD3-559333-2021]
  4. OpenBCI Sponsorship Program

Ask authors/readers for more resources

This study investigated the impact of sampling frequency on the validity and reliability of deriving cardiac metrics. Results showed that downsampling frequencies of >= 50 and >= 90 Hz produced highly robust measures, while lower frequencies resulted in wider variability and less accuracy. Further research is needed to determine the minimum sampling frequency required for obtaining valid heart rate and HRV metrics from pulsatile waveforms.
Background: While cardiac pulsations are widely present within physiological and neuroimaging data, it is unknown the extent this information can provide valid and reliable heart rate and heart rate variability (HRV) estimates. The objective of this study was to demonstrate how a slight temporal shift due to an insufficient sampling frequency can impact the validity/accuracy of deriving cardiac metrics. Methods: Twenty-two participants were instrumented with valid/reliable industry-standard or open-source electrocardiograms. Five-minute lead II recordings were collected at 1000 Hz in an upright orthostatic position. Following artifact removal, the 1000 Hz recording for each participant was downsampled to frequencies ranging 2-500 Hz. The validity of each participant's downsampled recording was compared against their 1000 Hz recording (reference-standard) using Bland-Altman plots with 95 % limits of agreement (LOA), coefficient of variation (CoV), intraclass correlation coefficients, and adjusted r-squared values. Results: Downsampled frequencies of >= 50 and >= 90 Hz produced highly robust measures with narrow logtransformed 95 % LOA (<+/- 0.01) and low CoV values (<= 3.5 %) for heart rate and HRV metrics, respectively. Below these thresholds, the log-transformed 95 % LOA became wider (LOA range: +/- 0.1-1.9) and more variable (CoV range: 1.5-111.6 %). Conclusion: These results provide an important consideration for obtaining cardiac information from physiological data. Compared to the reference-standard ECG, a seemingly negligible temporal shift of the systolic contraction (R wave) greater than 11-milliseconds (90 Hz) away from its true value, lessened the validity of the HRV. Further research is warranted to determine the minimum sampling frequency required to obtain valid heart rate/HRV metrics from pulsatile waveforms.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available