Strontium isotopic analysis of environmental microsamples by inductively coupled plasma - tandem mass spectrometry

Journal

JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY
Volume 37, Issue 1, Pages 103-113

Publisher

ROYAL SOC CHEMISTRY
DOI: 10.1039/d1ja00329a

Keywords

-

Funding

  1. PNRA (Programma Nazionale di Ricerca in Antartide) [PNRA 2016/AC2.04 (PNRA 16_00252-Linea A2), PNRA 2015/AC3.04 (PNRA 14_00091-Linea A3)]
  2. FWO-Vlaanderen [12ZA320N]
  3. MIUR (Italian Ministry of University and Research)

Ask authors/readers for more resources

A new method has been developed for accurate and precise Sr isotopic analysis of microsamples via ICP-MS/MS, utilizing a multivariate approach to optimize instrument settings for maximum sensitivity and precision of isotope ratios. The method was found to be accurate and precise, with external precision of 0.17% at an analytical concentration of 2 ng g(-1). Results showed good agreement with literature data for samples with low Sr concentrations, demonstrating the suitability of the procedure for complex matrices with limited sample amounts.
In this work, a new method has been developed for accurate and precise Sr isotopic analysis of microsamples via inductively coupled plasma - tandem mass spectrometry (ICP-MS/MS). For introduction of low-volume samples, a setup consisting of a syringe-driven pump that allows to work at low and stable sample introduction flow rates was combined with a high-efficiency sample introduction system, originally designed for the introduction of single cells and other discrete entities into the ICP. The method requires 240 mu L of sample only, introduced at a sample uptake rate of 20 mu L min(-1) and allows for the straightforward measurement of the Sr-87/Sr-86 isotope ratio without prior Rb/Sr separation. The isobaric overlap of the signals of Rb-87 and Sr-87 is overcome by the selective reaction between Sr+ and CH3F gas in the collision/reaction cell (CRC) of an ICP-MS/MS instrument. In this way, Sr+ ions are measured as their corresponding SrF+ reaction product ions in a mass-shift approach. A multivariate approach was applied to obtain optimum instrument settings (sample uptake rate, nebulizer gas flow rate, make-up gas flow rate, and plasma sampling depth) for maximum sensitivity and optimum precision of the Sr-87/Sr-86 and Sr-88/Sr-86 isotope ratios. The method thus developed was found to be accurate (mean bias = -0.001 +/- 0.070%, with respect to a quality control reference value) and showed an external precision of 0.17% (RSD, n = 24) at an analytical concentration of 2 ng g(-1). The method was applied to the analysis of small-volume digests of samples of Antarctic atmospheric particulates (PM10) and of soils from Australia and South America as potential source areas for atmospheric particulates reaching East Antarctica. The raw soil samples were handled using a dedicated resuspension/collection system to select the finest fraction (PM2.5), thus simulating the long-range transport of mineral dust in the atmosphere. The results obtained for these samples were found to be in good agreement with literature data, demonstrating the suitability of the procedure developed for the Sr isotopic analysis of such complex matrices with Sr concentrations as low as a few ng g(-1) (<3 ng g(-1) and <7 ng g(-1), for the PM10 and soil samples, respectively). The method could also be useful for other environmental samples for which the low Sr concentrations and limited sample amounts are critical aspects.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available