4.7 Article

Evaluation of poly(styrene-d5) and poly(4-fluorostyrene) as internal standards for microplastics quantification by thermoanalytical methods

Journal

Publisher

ELSEVIER
DOI: 10.1016/j.jaap.2021.105310

Keywords

Internal standard; Microplastics; Poly(4-fluorostyrene); Poly(styrene-d5)

Funding

  1. Federal Ministry of Transport and Digital Infrastructure (BMVI)
  2. University Koblenz-Landau
  3. German Environment Agency (UBA)

Ask authors/readers for more resources

Thermoanalytical methods like Py-GC-MS are promising for microplastics quantification in environmental samples, but lack harmonization and standardization. The use of an internal standard can improve data quality and method robustness. Poly(4fluorostyrene) shows potential as an alternative to poly(styrene-d5) in quantification, with no exchange reactions during pyrolysis.
Thermoanalytical methods such as pyrolysis-gas chromatography-mass spectrometry (Py-GC-MS) are among the most promising techniques for the quantification of microplastics (MP) in environmental samples. However, methods still lack harmonization and standardization. The use of an internal standard (IS) can improve the quality of quantitative data and the robustness of the analytical method. Poly(styrene-d5) has frequently been used as IS for MP quantification, but suffers from H-D exchange during pyrolysis, which is catalyzed by inorganic components of the sample matrix, most evidently in aluminum oxide filter matrix and in sea sand. Poly(4fluorostyrene) (PFS) is a promising alternative, which does not suffer from exchange reactions. Both standards yielded comparable results in calibrations for polypropylene, polyethylene, and polystyrene MP, but a larger concentration range could be covered if PFS was used. Only minor interferences of both polymers with natural organic matrix components were observed. The study of influences of various inorganic matrices on pyrolysis of the polymers revealed manifold effects that may interfere with MP quantification and underline the need for matrix-matched calibration or extensive sample clean-up.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available