PET Timing Performance Measurement Method Using NEMA NEC Phantom

Journal

IEEE TRANSACTIONS ON NUCLEAR SCIENCE
Volume 63, Issue 3, Pages 1335-1342

Publisher

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TNS.2016.2522766

Keywords

Biomedical imaging; molecular imaging; nuclear medicine; positron emission tomography; tomography; whole-body PET

Ask authors/readers for more resources

When comparing the performance of time-of-flight whole-body PET scanners, timing resolution is one important benchmark. Timing performance is heavily influenced by detector and electronics design. Even for the same scanner design, measured timing resolution is a function of many factors including the activity concentration, geometry and positioning of the radioactive source. Due to lack of measurement standards, the timing resolutions reported in the literature may not be directly comparable and may not describe the timing performance under clinically relevant conditions. In this work we introduce a method which makes use of the data acquired during the standard NEMA Noise-Equivalent-Count-Rate (NECR) measurements, and compare it to several other timing resolution measurement methods. The use of the NEMA NEC phantom, with well-defined dimensions and radioactivity distribution, is attractive because it has been widely accepted in the industry and allows for the characterization of timing resolution across a more relevant range of conditions.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available