4.5 Article

Evaluation of the Visby medical COVID-19 point of care nucleic acid amplification test

Journal

CLINICAL BIOCHEMISTRY
Volume 117, Issue -, Pages -

Publisher

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.clinbiochem.2021.11.007

Keywords

Severe acute respiratory syndrome coronavirus; 2 (SARS-CoV-2); Nucleic acid amplification test (NAAT); Point of care (POC); Coronavirus disease 2019 (COVID-19); Rapid test

Ask authors/readers for more resources

The Visby Medical COVID-19 test is a rapid and accurate method for detecting SARS-CoV-2. It has high concordance with laboratory-based reference methods and superior sensitivity compared to other point-of-care devices.
Rapid and widespread diagnostic testing is critical to providing timely patient care and reducing transmission of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2). Recently, the Visby Medical COVID-19 point of care (POC) test was granted emergency use authorization (EUA) for qualitative detection of SARS-CoV-2 nucleic acid at the point of care. We evaluated its performance characteristics using residual specimens (n = 100) collected from Mayo Clinic patients using nasopharyngeal (NP) swabs and placed in viral transport media (VTM). The same specimen was tested using both the laboratory reference method (RT-qPCR) and Visby test. The reference methods utilized included a laboratory developed test with EUA (Mayo Clinic Laboratories, Rochester, MN) using the TaqMan assay on a Roche Light Cycler 480 or a commercially available EUA platform (cobas (R) SARS-CoV-2; Roche Diagnostics, Indianapolis, IN). Positive, negative, and overall percent agreement between the Visby COVID-19 test and the reference method were calculated. Additionally, the limit of detection (LoD) claimed by the manufacturer (1112 copies/mL) was verified with serial dilutions of heat inactivated virus. The Visby COVID-19 test correctly identified 29/30 positive samples and 69/70 negative samples, resulting in an overall concordance of 98.0%, positive percent agreement of 96.7%, and negative percent agreement of 98.6%. The abbreviated LoD experiment showed that the analytical sensitivity of the method is as low as or lower than 500 copies/mL. Our study demonstrated that Visby COVID-19 is well-suited to address rapid SARS-CoV-2 testing needs. It has high concordance with central laboratory-based RT-qPCR methods, a low rate of invalid results, and superior analytical sensitivity to some other EUA POC devices.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available