4.5 Article

Vitamin E infused highly cross-linked cemented cups in total hip arthroplasty show good wear pattern and stabilize satisfactorily: a randomized, controlled RSA trial with 5-year follow-up

Journal

ACTA ORTHOPAEDICA
Volume 93, Issue -, Pages 249-255

Publisher

Medical Journal Sweden AB
DOI: 10.2340/17453674.2022.1517

Keywords

-

Categories

Ask authors/readers for more resources

This study compared the wear characteristics and clinical outcomes of vitamin E infused highly cross-linked polyethylene cups with regular ultra-high molecular weight polyethylene cups. The results showed reduced wear and good clinical outcomes in the vitamin E infused cups up to 5 years.
Background and purpose - Vitamin E infused highly cross-linked polyethylene (VEPE) was introduced in order to enhance oxidative resistance in highly cross-linked polyethylene cups in total hip arthroplasty (THA). We have, with a follow-up of 5 years, evaluated wear characteristics of 2 identically designed cemented cups with the only difference being the material, VEPE or ultra-high molecular weight polyethylene (UHMWPE). Furthermore, we report cup migration and clinical outcome. Patients and methods - 48 patients with primary osteoarthritis were randomized to either UHMWPE or VEPE cups. Patients were followed with radiostereometric analysis (RSA) from the first postoperative day, at 3 months, 1, 2, and 5 years as well as with hip-specific outcome questionnaires. Results - At 3 months the mean proximal head penetration for UHMWPE was 0.07 mm (95% CI 0.03-0.11) and for VEPE 0.06 mm (-0.01 to 0.13). Thereafter, there was a continuous annual wear of 0.08 mm/year, up to 0.46 mm (0.36-0.57) at 5 years, for the UHMWPE cup. The VEPE cup showed low annual wear of 0.01 mm/year, up to 0.09 mm (0.02-0.16) at 5 years. In the first 3 months the UHMWPE cup migrated cranially 0.08 mm (0.03-0.13) whereas the VEPE cup migrated 0.17 mm (0.10-0.24), Thereafter, they showed similar migration patterns with stabilization between 2 and 5 years up to 0.21 mm (0.04-0.39) and 0.24 mm (0.13-0.36) respectively. The HOOS remained good up to 5 years, and no cup was revised. Interpretation - Compared with otherwise identical UHMWPE cups the cemented VEPE cup shows statistically significant reduction of wear up to 5 years and both cup types stabilize well with good clinical outcome.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available