Voxel-based partial volume correction of PET images via subtle MRI guided non-local means regularization

Journal

Publisher

ELSEVIER SCI LTD
DOI: 10.1016/j.ejmp.2021.07.028

Keywords

PET; Partial Volume Correction; NLMA; Non-Local Means; Anatomical guidance

Funding

  1. Natural Science Foundation of Guangdong Province [2018A030313366]
  2. National Institutes of Health [AG056142]

Ask authors/readers for more resources

This study introduces a partial volume correction (PVC) method using anatomical non-local means (NLMA) regularization to improve PET image quality. By incorporating MRI information and optimizing the LS objective function, better visual and quantitative performance improvements were achieved.
Purpose: Positron emission tomography (PET) images tend to be significantly degraded by the partial volume effect (PVE) resulting from the limited spatial resolution of the reconstructed images. Our purpose is to propose a partial volume correction (PVC) method to tackle this issue. Methods: In the present work, we explore a voxel-based PVC method under the least squares framework (LS) employing anatomical non-local means (NLMA) regularization. The well-known non-local means (NLM) filter utilizes the high degree of information redundancy that typically exists in images, and is typically used to directly reduce image noise by replacing each voxel intensity with a weighted average of its non-local neighbors. Here we explore NLM as a regularization term within iterative-deconvolution model to perform PVC. Further, an anatomical-guided version of NLM was proposed that incorporates MRI information into NLM to improve resolution and suppress image noise. The proposed approach makes subtle usage of the accompanying MRI information to define a more appropriate search space within the prior model. To optimize the regularized LS objective function, we used the Gauss-Seidel (GS) algorithm with the one-step-late (OSL) technique. Results: After the import of NLMA, the visual and quality results are all improved. With a visual check, we notice that NLMA reduce the noise compared to other PVC methods. This is also validated in bias-noise curve compared to non-MRI-guided PVC framework. We can see NLMA gives better bias-noise trade-off compared to other PVC methods. Conclusions: Our efforts were evaluated in the base of amyloid brain PET imaging using the BrainWeb phantom and in vivo human data. We also compared our method with other PVC methods. Overall, we demonstrated the value of introducing subtle MRI-guidance in the regularization process, the proposed NLMA method resulting in promising visual as well as quantitative performance improvements.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.4
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available