4.1 Article

Comparative efficacy of C-MAC Miller videolaryngoscope versus McGrath MAC size 1 videolaryngoscope in neonates and infants undergoing surgical procedures under general anesthesia: A prospective randomized controlled trial

Journal

PEDIATRIC ANESTHESIA
Volume 31, Issue 10, Pages 1089-1096

Publisher

WILEY
DOI: 10.1111/pan.14244

Keywords

glottis; infant; intubation; laryngoscopy; newborn

Ask authors/readers for more resources

The study found that C-MAC Miller and McGrath MAC showed similar intubation timings, success rates, and intubation difficulty scores in neonates and infants, but the former provided superior glottic views.
Background Various anatomical and physiological factors make intubation in infants challenging. C-MAC videolaryngoscope shows better results as compared to the conventional direct laryngoscopy for intubation in infants. McGrath MAC size-1 with a disposable Macintosh type blade has recently been introduced for use in infants and has not been formally evaluated in this population. Aims This study aims to evaluate the intubation characteristics of C-MAC Miller and McGrath MAC in neonates and infants with the primary objective to compare the time with the two devices. Methods After informed consent from the parents, 140 neonates and infants scheduled for surgical procedures were randomized to undergo intubation with either C-MAC Miller or McGrath MAC after standard general anesthesia. The two devices were compared in terms of total intubation time, Percent of Glottic Opening score, Cormack Lehane grades, time to glottis view, intubation difficulty score, overall success rate, first attempt success rate, and complications. Results The median glottic view time (6 s [4-9] vs. 6 s [4-9]; p = .40) and intubation time (27 s [25.5-28] vs. 27 s [24.5-29.5]; p = .87) were similar. The mean difference (95% CI) in time to tracheal intubation and time to glottic view was 0.49 s [-3.1 to 2.1] and -1.7 s [-3.8 to 0.47], respectively. However, the Percent of Glottic Opening score, Cormack Lehane grades, and subjective intubation difficulty were significantly better with C-MAC. The first attempt success rates, overall success rates (100% vs. 97.5%), and intubation difficulty scores were comparable. There were two failed intubations with McGrath which were successfully intubated with C-MAC. Conclusion The C-MAC Miller blade showed similar intubation timings, success rates, and intubation difficulty score as compared to McGrath MAC in neonates and infants, though the former provided superior glottic views. Both the videolaryngoscopes may be safely used in infants and neonates for routine intubation scenarios.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.1
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available