4.5 Article

Significant differences in rates of aseptic loosening between two variations of a popular total knee arthroplasty design

Journal

INTERNATIONAL ORTHOPAEDICS
Volume 45, Issue 11, Pages 2859-2867

Publisher

SPRINGER
DOI: 10.1007/s00264-021-05151-w

Keywords

Total knee arthroplasty; Aseptic loosening; NexGen; Revision; Knee

Categories

Ask authors/readers for more resources

This study found an increased risk of aseptic loosening in non-augmentable components, highlighting the impact of minor implant changes on outcomes. Clinicians are advised to remain vigilant to implant changes and publish their own research results when important trends are observed.
Purpose The NexGen Legacy Posterior Stabilised (LPS) prosthesis (Zimmer Biomet, Warsaw, IN, USA) has augmentable and non-augmentable tibial baseplate options. We have noted an anecdotal increase in the number of cases requiring early revision for aseptic loosening since adopting the non-augmentable option. The purpose of this study was to ascertain our rates of aseptic tibial loosening for the two implant types within five years of implantation and to investigate the causes for any difference observed. Methods A database search was performed for all patients who underwent primary total knee arthroplasty (TKA) using the NexGen LPS between 2009 and 2015. Kaplan-Meier curves were plotted to assess for differences in revision rates between cohorts. We collected and compared data on gender, age, body mass index, component alignment and cement mantle quality as these were factors thought to affect the likelihood of aseptic loosening. Results Two thousand one hundred seventy-two TKAs were included with five year follow-up. There were 759 augmentable knees of which 14 were revised and 1413 non-augmentable knees of which 48 were revised. The overall revision rate at five years was 1.84% in the augmentable cohort and 3.4% in the non-augmentable cohort. The revision rate for aseptic loosening was 0.26% in the augmentable group and 1.42% in the non-augmentable group (p = 0.0241). Conclusions We have identified increased rates of aseptic loosening in non-augmentable components. This highlights the effect that minor implant changes can have on outcomes. We recommend that clinicians remain alert to implant changes and publish their own results when important trends are observed.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available