4.7 Article

Measurement of the Lag Correction Factor in Low-Dose Fluoroscopic Imaging

Journal

IEEE TRANSACTIONS ON MEDICAL IMAGING
Volume 40, Issue 6, Pages 1661-1672

Publisher

IEEE-INST ELECTRICAL ELECTRONICS ENGINEERS INC
DOI: 10.1109/TMI.2021.3063350

Keywords

Detectors; X-rays; Logic gates; Heuristic algorithms; Thin film transistors; Electron tubes; Imaging; Exposure leak; flat-panel dynamic detector; fluoroscopic imaging; Granfors-Aufrichtig method; lag correction factor

Funding

  1. National Research Foundation of Korea (NRF) - Korea Government (MISP) [2020R1A2C1009895]
  2. Hankuk University of Foreign Studies Research Fund of 2021
  3. National Research Foundation of Korea [2020R1A2C1009895] Funding Source: Korea Institute of Science & Technology Information (KISTI), National Science & Technology Information Service (NTIS)

Ask authors/readers for more resources

This study focuses on the lag signals in flat-panel dynamic detectors, proposing the need to accurately evaluate detector lag through quantitative measurement and introducing the use of lag correction factors. Additionally, the study compares the effectiveness of two different measurement methods.
Lag signals occur at images sequentially acquired from a flat-panel (FP) dynamic detector in fluoroscopic imaging due to charge trapping in photodiodes and incomplete readouts. This lag signal produces various lag artifacts and prevents analyzing detector performances because the measured noise power spectrum (NPS) values are reduced. In order to design dynamic detectors, which produce low lag artifacts, accurately evaluating the detector lag through its quantitative measurement is required. A lag correction factor can be used to both examine the detector lag and correct measured NPS. To measure the lag correction factor, the standard of IEC62220-1-3 suggests a temporal power spectral density under a constant potential generator for the x-rays. However, this approach is sensitive to disturbing noise and thus becomes a problem in obtaining accurate estimates especially at low doses. The Granfors-Aufrichtig (GA) method is appropriate for noisy environments with a synchronized pulse x-ray source. However, for the x-ray source of a constant potential generator, gate-line scanning to read out charges produces a nonuniform lag signal within each image frame and thus the conventional GA method yields wrong estimates. In this paper, we first analyze the GA method and show that the method is an asymptotically unbiased estimate. Based on the GA method, we then propose three algorithms considering the scanning process and exposure leak, in which line estimates along the gate line are exploited. We extensively conducted experiments for FP dynamic detectors and compared the results with conventional algorithms.