4.4 Article

Long-term complications in patients implanted with subcutaneous implantable cardioverter-defibrillators: Real-world data from the extended ELISIR experience

Journal

HEART RHYTHM
Volume 18, Issue 12, Pages 2050-2058

Publisher

ELSEVIER SCIENCE INC
DOI: 10.1016/j.hrthm.2021.07.008

Keywords

ICD complications; Lead complications; S-ICD; S-ICD recall; Subcutaneous implantable cardioverter-defibrillator

Ask authors/readers for more resources

The study aimed to evaluate device-related complications in a large real-world multicenter S-ICD cohort. Over 23.2 months of follow-up, the overall complication rate was 9.3%. Early unanticipated device battery depletions occurred in 2.2% of patients, while lead fracture was observed in 0.3%.
BACKGROUND Recently, the Food and Drug Administration issued a recall for the subcutaneous implantable cardioverter-defibrillator (S-ICD) because of the possibility of lead ruptures and accelerated battery depletion. OBJECTIVE The aim of this study was to evaluate device-related complications over time in a large real-world multicenter S-ICD cohort. METHODS Patients implanted with an S-ICD from January 2015 to June 2020 were enrolled from a 19-institution European registry (Experience from the Long-term Italian S- ICD registry [ELISIR]; ClinicalTrials.gov identifier NCT0473876). Device-related complication rates over follow-up were collected. Last follow-up of patients was performed after the Boston Scientific recall issue. RESULTS A total of 1254 patients (median age 52.0 [interquartile range 41.0-62.2] years; 973 (77.6%) men; 387 (30.9%) ischemic) was enrolled. Over a follow-up of 23.2 (12.8-37.8) months, complications were observed in 117 patients (9.3%) for a total of 127 device-related complications (23.6% managed conservatively and 76.4% required reintervention). Twenty-seven patients (2.2%) had unanticipated generator replacement after 3.6 (3.3-3.9) years, while 4 (0.3%) had lead rupture. Body mass index (hazard ratio [HR] 1.063 [95% confidence interval 1.028-1.100]; P<.001), chronic kidney disease (HR 1.960 [1.191-3.225]; P=.008), and oral anticoagulation (HR 1.437 [1.010-2.045]; P =.043) were associated with an increase in overall complications, whereas older age (HR 0.980 [0.967-0.994]; P=.007) and procedure performed in high-volume centers (HR 0.463 [0.300-0.715]; P=.001) were protective factors. CONCLUSION The overall complication rate over 23.2 months of follow-up in a multicenter S-ICD cohort was 9.3%. Early unanticipated device battery depletions occurred in 2.2% of patients, while lead fracture was observed in 0.3%, which is in line with the expected rates reported by Boston Scientific.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.4
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available