4.4 Article

Altered Media Flow and Tablet Position as Factors of How Air Bubbles Affect Dissolution of Disintegrating and Non-disintegrating Tablets Using a USP 4 Flow-Through Cell Apparatus

Journal

AAPS PHARMSCITECH
Volume 22, Issue 7, Pages -

Publisher

SPRINGER
DOI: 10.1208/s12249-021-02117-4

Keywords

flow-through cell dissolution test; dissolved gas; fluid velocity; tablet position

Funding

  1. Research on Regulatory Harmonization and Evaluation of Pharmaceuticals
  2. Japan Agency for Medical Research and Development (AMED) [JP19mk0101130]

Ask authors/readers for more resources

This study used PIV and CFD techniques to investigate the impact of air bubbles on tablet dissolution in a flow-through cell system, finding that bubbles can alter local flow dynamics, affecting dissolution rates, and highlighting the importance of deaeration in dissolution studies.
This study investigated how air bubbles in media affect tablet dissolution in a flow-through cell system (USP 4) using disintegrating (USP prednisone) and non-disintegrating (USP salicylic acid) tablets. Cell hydrodynamics were studied using particle image velocimetry (PIV) and computational fluid dynamics (CFD). The PIV analysis showed periodic changes in the local flow corresponding to the discharge and suction of the pump cycles. The absence of prior deaeration induced small air bubbles in the media and lower maximum flow during the cycle, explaining the slower dissolution of the USP salicylic acid tablets. Bubbles, occurring during the USP prednisone tablets study, induced the transition of floating disintegrated particles towards the cell outlet, whereas the particles precipitated to form a white layer on the glass beads used in the study with prior deaeration. CFD analysis showed local flow variation in multiple positions of small (ID 12 mm) and large (ID 22.6 mm) cells, explaining the different rates of dissolution of prednisone tablet particles depending on their distribution. These results emphasize the importance of prior deaeration in dissolution studies using a flow-through system. Bubbles in the flow-through cell system affected tablet dissolution by reducing the area in contact with the media (wettability), lowering the maximum instantaneous flow (pressure buffering), and altering the position of disintegrated particles in the cell.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.4
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available