4.6 Article

Inactivating SARS-CoV-2 Using 275 nm UV-C LEDs through a Spherical Irradiation Box: Design, Characterization and Validation

Journal

MATERIALS
Volume 14, Issue 9, Pages -

Publisher

MDPI
DOI: 10.3390/ma14092315

Keywords

SARS-CoV-2; COVID-19; LED; UVC; disinfection; inactivation; virucide

Ask authors/readers for more resources

A spherical irradiation system based on UV-C 275 nm LEDs was designed, characterized and validated for inactivating SARS-CoV-2. Results showed that a 99.9% inactivation rate can be achieved after 1 minute of treatment with a dose of 83.1 J/m(2).
We report on the design, characterization and validation of a spherical irradiation system for inactivating SARS-CoV-2, based on UV-C 275 nm LEDs. The system is designed to maximize irradiation intensity and uniformity and can be used for irradiating a volume of 18 L. To this aim: (i) several commercially available LEDs have been acquired and analyzed; (ii) a complete optical study has been carried out in order to optimize the efficacy of the system; (iii) the resulting prototype has been characterized optically and tested for the inactivation of SARS-CoV-2 for different exposure times, doses and surface types; (iv) the result achieved and the efficacy of the prototype have been compared with similar devices based on different technologies. Results indicate that a 99.9% inactivation can be reached after 1 min of treatment with a dose of 83.1 J/m(2).

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available