4.2 Article

Heater-Cooler Devices and Risk of Contamination during Cardiac Surgery

Journal

THORACIC AND CARDIOVASCULAR SURGEON
Volume 69, Issue 8, Pages 703-709

Publisher

GEORG THIEME VERLAG KG
DOI: 10.1055/s-0040-1721745

Keywords

cardiopulmonary bypass; CPB; wound infection

Ask authors/readers for more resources

The study investigated microbial contamination in new Heater-cooler devices (HCDs) used in cardiothoracic surgery, finding that despite decontamination procedures, microbial growth was identified in all devices over time. However, airborne transmission of microorganisms directly from the HCD to the operating field appears unlikely.
Background Heater-cooler devices (HCD) have been implicated in a cardiosurgical contamination scenario causing prosthetic valve endocarditis. Aim We characterized contamination of new HCDs and assessed the risk of intraoperative microorganism transmission from the HCD to the operating field. Methods We initially acquired four new FlexTherm and then four new Maquet HCU40 HCDs and assessed occurrence and speed of microbial contamination (including mycobacteria) assessing swab and water samples from the device. In parallel, we collected repeated samples from different sites in the operating room either by swab sticks or by exposing different sample plates to room air. We also reviewed microbiological results from the hospital and compared them to cardiosurgical wound infections and endocarditis cases. Finally, we simulated cardiosurgical conditions and assessed the devices' ability to expel air to the operative field. Results All new HCDs were clean before first use. Despite authority-mandated decontamination procedures, microbial growth ( Fusarium solani , Sphingomonas paucimobilis , Pseudomonas aeruginosa , Mycobacterium chelonae , and gordonae ) was identified in all HCDs over time and could not be permanently eliminated. Four of these mircoorganisms were also found in tap water. However, none of the HCD-organisms were found inside the laminar airflow operating area. Importantly, except for P. aeruginosa , none of the HCD organisms were found in patients with surgical wound infections or endocarditis. HCD-expelled air did not rise more than 40cm above ground. Conclusion HCDs cannot be expected to remain sterile despite extensive decontamination procedures. However, airborne transmission of microorganisms directly from the HCD to the operating field appears unlikely.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.2
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available