4.6 Article

Design and Characterization of Electrochemical Sensor for the Determination of Mercury(II) Ion in Real Samples Based upon a New Schiff Base Derivative as an Ionophore

Journal

SENSORS
Volume 21, Issue 9, Pages -

Publisher

MDPI
DOI: 10.3390/s21093020

Keywords

Schiff base; mercury selective electrode; ionophore; PVC membrane; ISE-Hg

Funding

  1. Taif University

Ask authors/readers for more resources

This study presents a new Hg2+ selective electrode with fast response time and low interference for detecting ultra-trace levels of Hg2+ in various real samples. The electrode shows a linear response to Hg2+ ions in a wide concentration range, with a low detection limit. The developed electrode was compared with ICP-OES measurements for accuracy and was found to perform well.
The present paper provides a description of the design, characterization, and use of a Hg2+ selective electrode (Hg2+-SE) for the determination of Hg2+ at ultra-traces levels in a variety of real samples. The ionophore in the proposed electrode is a new Schiff base, namely 4-bromo-2-[(4-methoxyphenylimino)methyl]phenol (BMPMP). All factors affecting electrode response including polymeric membrane composition, concentration of internal solution, pH sample solution, and response time were optimized. The optimum response of our electrode was obtained with the following polymeric membrane composition (% w/w): PVC, 32; o-NPOE, 64.5; BMPMP, 2 and NaTPB, 1.5. The potentiometric response of Hg2+-SE towards Hg2+ ion was linear in the wide range of concentrations (9.33 x 10(-8)-3.98 x 10(-3) molL(-1)), while, the limit of detection of the proposed electrode was 3.98 x 10(-8) molL(-1) (8.00 mu g L-1). The Hg2+-SE responds quickly to Hg2+ ions as the response time of less than 10 s. On the other hand, the slope value obtained for the developed electrode was 29.74 +/- 0.1 mV/decade in the pH range of 2.0-9.0 in good agreement with the Nernstian response (29.50 mV/decade). The Hg2+-SE has relatively less interference with other metal ions. The Hg2+-SE was used as an indicator electrode in potentiometric titrations to estimate Hg2+ ions in waters, compact fluorescent lamp, and dental amalgam alloy and the accuracy of the developed electrode was compared with ICP-OES measurement values. Moreover, the new Schiff base (BMPMP) was synthesized and characterized using ATR-FTIR, elemental analysis, H-1 NMR, and C-13 NMR. The PVC membranes containing BMPMP as an ionophore unloaded and loaded with Hg(II) are reported by scanning electron microscope images (SEM) along with energy-dispersive X-ray spectroscopy (EDX) spectra.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available