Establishment of rapid, sensitive, and quantitative liquid chromatography-electrospray ionization-tandem mass spectrometry method coupled with liquid-liquid extraction for measurement of urinary bisphenol A, 4-t-octylphenol, and 4-nonylphenol

Journal

RAPID COMMUNICATIONS IN MASS SPECTROMETRY
Volume 35, Issue 12, Pages -

Publisher

WILEY
DOI: 10.1002/rcm.9084

Keywords

-

Funding

  1. Mersin University, Department of Scientific Research Projects [BAP-ECZ F EMBB [DB] 2012-8 A]

Ask authors/readers for more resources

A method for the trace analysis of endocrine disruptors such as Bisphenol A, 4-t-octylphenol, and 4-nonylphenol in human urine using liquid chromatography-mass spectrometry was developed and shown to be accurate and sensitive. The method was successfully applied for the determination of these compounds in 15 human urine samples, demonstrating its effectiveness in detecting these chemicals.
Rationale Bisphenol A (BPA), 4-t-octylphenol (4-t-OP), and 4-nonylphenol (4-NP) are ubiquitous manufactured chemicals classified as endocrine disruptors. Now these substances are prevalent and found in a wide range of biological and environmental matrices globally. Acting as xenoestrogens, their target is to bind estrogen receptors and compete against natural hormones. Thus, there is an increasing concern on their negative health consequences. Methods A rapid, sensitive, and robust method was established, validated, and used for the trace analysis of BPA, 4-t-OP, and 4-NP in human urine using liquid chromatography-electrospray ionization-tandem mass spectrometry. Results The linear ranges for BPA, 4-t-OP, and 4-NP were 5.0-500 ng/mL. The correlation coefficient was higher than 0.996. The limit of detection and limit of quantitation for the compounds were 0.001, 0.007, and 0.005 ng/mL and 0.005, 0.023, and 0.180 ng/mL, respectively. The recoveries were as follows: 99.539%-108.557%, 97.780%-110.768%, and 99.694%-106.908% for BPA, 4-NP, and 4-t-OP respectively, with relative standard deviation The established method was also applied effectively for the determination of BPA, 4-t-OP, and 4-NP levels in 15 human urine samples. This method was specifically developed for easy and precise analysis of BPA, 4-t-OP, and 4-NP in human urine. The method developed has been shown to be accurate, precise, and sensitive.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.4
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available