4.6 Article

Shape accuracy of fiber optic sensing for medical devices characterized in bench experiments

Journal

MEDICAL PHYSICS
Volume 48, Issue 7, Pages 3936-3947

Publisher

WILEY
DOI: 10.1002/mp.14881

Keywords

distributed optical fiber strain sensing; navigation; optical frequency domain reflectometry; radiation free; real time

Ask authors/readers for more resources

The FORS technology accurately visualizes the shape of medical devices in three dimensions, providing guidance and navigation during minimally invasive procedures. Bench experiments demonstrated submillimeter accuracy in tip location and milliradian accuracy in orientation. The technology surpasses other navigation and tracking technologies in precision, potentially reducing the need for fluoroscopy during procedures.
Purpose Fiber Optic RealShape (FORS) is a new technology that visualizes the full three-dimensional shape of medical devices, such as catheters and guidewires, using an optical fiber embedded in the device. This three-dimensional shape provides guidance to clinicians during minimally invasive procedures, and enables intuitive navigation. The objective of this paper is to assess the accuracy of the FORS technology, as implemented in the current state-of-the-art Philips FORS system. The FORS system provides the shape of the entire device, including tip location and orientation. We consider all three aspects. Methods In bench experiments, we determined the accuracy of the location and orientation of the tip by displacing and rotating the fiber end, while allowing the rest of the fiber to change shape freely. To test the accuracy of the full shape, we have placed the fiber in a groove, which was accurately machined in a thick, stiff metal path plate. We then compared the reconstructed shape with the known shape of the groove. Results The tip location is found with submillimeter accuracy, and the orientation is sensed with milliradian accuracy. The shape of a fiber in the path plate faithfully follows the known shape of the groove, with typical deviation less than 0.5 mm in the plane of the plate. Out of plane accuracy, perhaps slightly less relevant clinically, is more challenging, due to the influence of twist; yet even out of the plane, the deviation is only submillimeter. Conclusion The technology achieves submillimeter precision and provides full three-dimensional shape, surpassing the reported precision of other navigation and tracking technologies, and therefore may potentially alleviate the need for fluoroscopy.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available