4.0 Article

In-vitro performance of a single-chambered total artificial heart in a Fontan circulation

Journal

JOURNAL OF ARTIFICIAL ORGANS
Volume 25, Issue 1, Pages 1-8

Publisher

SPRINGER JAPAN KK
DOI: 10.1007/s10047-021-01273-5

Keywords

Fontan circulation; Mock circulation loop (MCL); Total artificial heart (TAH); Single ventricle; Congenital heart disease

Funding

  1. Projekt DEAL

Ask authors/readers for more resources

An in-vitro study investigated the feasibility of using a single pumping chamber of the SynCardia total artificial heart as a replacement for a single ventricle palliated by Fontan circulation. The study showed that the single-chambered TAH performed sufficiently in the setting of the Fontan circulation, even achieving higher cardiac outputs in scenarios of pulmonary hypertension. This suggests that the replacement of a failing univentricular heart by a single chamber of the SynCardia TAH may provide a successful alternative for Fontan patients.
An in-vitro study was conducted to investigate the general feasibility of using only one pumping chamber of the SynCardia total artificial heart (TAH) as a replacement of the single ventricle palliated by Fontan circulation. A mock circulation loop was used to mimic a Fontan circulation. The combination of both ventricle sizes (50 and 70 cc) and driver (Freedom Driver and Companion C2 Driver) was investigated. Two clinical relevant scenarios (early Fontan; late Fontan) as derived from literature data were set up in the mock loop. The impact of increased transpulmonary pressure gradient, low atrial pressure, and raised central venous pressure on cardiac output was studied. From a hemodynamic point, the single-chambered TAH performed sufficiently in the setting of the Fontan circulation. Increased transpulmonary pressure gradient, from ideal to pulmonary hypertension, decreased the blood flow in combinations by almost 2 L/min. In the early Fontan scenario, a cardiac output of 3-3.5 L/min was achieved using the 50 cc ventricle, driven by the Companion C2 Driver. Even under pulmonary hypertension, cardiac outputs greater than 4 L/min could be obtained with the 70 cc pump chamber in the late Fontan scenario. In the clinically relevant Fontan scenarios, implementation of the single chambered TAH performed successfully from a hemodynamic point of view. The replacement of the failing univentricular heart by a single chamber of the SynCardia TAH may provide an alternative to a complex biventricular repair procedure or ventricular support in Fontan patients.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.0
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available