4.6 Article

Towards 'automated gonioscopy': a deep learning algorithm for 360° angle assessment by swept-source optical coherence tomography

Journal

BRITISH JOURNAL OF OPHTHALMOLOGY
Volume 106, Issue 10, Pages 1387-1392

Publisher

BMJ PUBLISHING GROUP
DOI: 10.1136/bjophthalmol-2020-318275

Keywords

angle; glaucoma; imaging

Categories

Funding

  1. National Medical Research Council and Biomedical Research Council, Singapore [10/1/35/19/674]

Ask authors/readers for more resources

The study validated a deep learning algorithm for 360 degrees angle assessment on swept-source optical coherence tomography, showing good diagnostic performance for detection of gonioscopic angle closure. This algorithm could potentially serve as the foundation for a non-contact, fast, and reproducible 'automated gonioscopy' in the future.
Aims To validate a deep learning (DL) algorithm (DLA) for 360 degrees angle assessment on swept-source optical coherence tomography (SS-OCT) (CASIA SS-1000, Tomey Corporation, Nagoya, Japan). Methods This was a reliability analysis from a cross-sectional study. An independent test set of 39 936 SS-OCT scans from 312 phakic subjects (128 SS-OCT meridional scans per eye) was analysed. Participants above 50 years with no previous history of intraocular surgery were consecutively recruited from glaucoma clinics. Indentation gonioscopy and dark room SS-OCT were performed. Gonioscopic angle closure was defined as non-visibility of the posterior trabecular meshwork in >= 180 degrees of the angle. For each subject, all images were analysed by a DL-based network based on the VGG-16 architecture, for gonioscopic angle-closure detection. Area under receiver operating characteristic curves (AUCs) and other diagnostic performance indicators were calculated for the DLA (index test) against gonioscopy (reference standard). Results Approximately 80% of the participants were Chinese, and more than half were women (57.4%). The prevalence of gonioscopic angle closure in this hospital-based sample was 20.2%. After analysing a total of 39 936 SS-OCT scans, the AUC of the DLA was 0.85 (95% CI:0.80 to 0.90, with sensitivity of 83% and a specificity of 87%) to classify gonioscopic angle closure with the optimal cut-off value of >35% of circumferential angle closure. Conclusions The DLA exhibited good diagnostic performance for detection of gonioscopic angle closure on 360 degrees SS-OCT scans in a glaucoma clinic setting. Such an algorithm, independent of the identification of the scleral spur, may be the foundation for a non-contact, fast and reproducible 'automated gonioscopy' in future.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.6
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available