4.4 Review

Endovascular Stapler Complications During Minimally Invasive Nephrectomy: An Updated Review of the FDA MAUDE Database From 2009-2019

Journal

UROLOGY
Volume 153, Issue -, Pages 181-184

Publisher

ELSEVIER SCIENCE INC
DOI: 10.1016/j.urology.2021.02.010

Keywords

-

Ask authors/readers for more resources

The study characterized the rates of endovascular stapler complications during minimally invasive radical nephrectomy over the last 10 years. Among the three types of staplers, TA staplers had a lower incidence of complications and were not associated with deaths or reoperations. Manufacturer evaluation suggested that human errors were a common cause of stapling malfunctions, emphasizing the importance of proper training for operating room staff to prevent potential complications.
OBJECTIVE To characterize the rates of endovascular stapler complications during hilar ligation in minimally invasive radical nephrectomy over the last 10 years. MATERIAL AND METHODS We reviewed the Food and Drug Administration Manufacturer and User Facility Device Experience database from January 1, 2009 to August 1, 2019. Staplers were categorized according to type, namely Ethicon Inc. endocutters (Johnson & Johnson, New Brunswick NJ); Endo-GIA (Medtronic, Minneapolis MN); and Endo-TA (Medtronic, Minneapolis MN). RESULTS There were 383 cases of complications involving staplers, 63% with Ethicon endocutters; 28% with GIA; and 9% with TA. 22 deaths (5.7% of total complications) were attributed to staplers. No deaths or reoperations occurred due to TA staplers. TA staplers were also associated with a reduced incidence of conversion to open as compared to Ethicon and GIA staplers. Apart from one device, manufacturer evaluation of returned devices either showed no abnormalities or attributed fault to improper use of staplers. DISCUSSION We characterized stapler complications during a 10-year period for minimally invasive radical nephrectomy. No deaths or reoperations occurred due to TA staplers, perhaps due to cutting and stapling occurring in separate steps. Based on manufacturer evaluation attributing stapling malfunctions to human errors, training of operating room staff on proper use of these devices is critical to prevent potentially significant complications from occurring. (C) 2021 Elsevier Inc.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.4
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available