Pre-validation of SENS-IS assay for in vitro skin sensitization of medical devices

Journal

TOXICOLOGY IN VITRO
Volume 71, Issue -, Pages -

Publisher

PERGAMON-ELSEVIER SCIENCE LTD
DOI: 10.1016/j.tiv.2020.105068

Keywords

-

Categories

Ask authors/readers for more resources

The study found that the SENS-IS assay can accurately identify sensitizers in medical device extracts, potentially replacing current in vivo assays.
According to ISO 10993-1:2018, the skin sensitization potential of all medical devices must be evaluated, and for this endpoint ISO 10993-10:2010 recommends the use of in vivo assays. The goal of the present study was to determine if the in vitro SENS-IS assay could be a suitable alternative to the current in vivo assays. The SENSIS assay uses the Episkin Large and SkinEthic RHE reconstructed human epidermis models to evaluate marker genes. In our study, the SENS-IS assay correctly identified 13 sensitizers spiked in a non-polar solvent. In a subsequent analysis six medical device silicone samples previously impregnated with sensitizers were extracted with polar and non-polar solvents. The SENS-IS assay correctly identified five of these extracts, while a sixth extract, which contained the weak sensitizer phenyl benzoate, was classified as negative. However, when this extract was concentrated, or a longer exposure time was used, the assay was able to detect phenyl benzoate. The SENS-IS assay was transferred to a naive laboratory which correctly identified sensitizers in six blinded silicone samples, including the one containing phenyl benzoate. In light of these results, we conclude that the SENS-IS assay is able to correctly identify the presence of sensitizers in medical devices extracts.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available