Predictors of Time-in-Range (70-180 mg/dL) Achieved Using a Closed-Loop Control System

Journal

DIABETES TECHNOLOGY & THERAPEUTICS
Volume 23, Issue 7, Pages 475-481

Publisher

MARY ANN LIEBERT, INC
DOI: 10.1089/dia.2020.0646

Keywords

Closed-loop systems; Type 1 diabetes; Insulin pump; Time-in-range

Funding

  1. Tandem Diabetes, Inc.
  2. National Institute of Diabetes and Digestive and Kidney Diseases

Ask authors/readers for more resources

Higher baseline TIR is the strongest predictor of TIR in children with T1D using closed-loop control systems, but lower baseline TIR is associated with the greatest improvement in TIR. User engagement is important for optimal glycemic control in closed-loop systems.
Background: Studies of closed-loop control (CLC) in patients with type 1 diabetes (T1D) consistently demonstrate improvements in glycemic control as measured by increased time-in-range (TIR) 70-180 mg/dL. However, clinical predictors of TIR in users of CLC systems are needed. Materials and Methods: We analyzed data from 100 children aged 6-13 years with T1D using the Tandem Control-IQ CLC system during a randomized trial or subsequent extension phase. Continuous glucose monitor data were collected at baseline and during 12-16 weeks of CLC use. Participants were stratified into quartiles of TIR on CLC to compare clinical characteristics. Results: TIR for those in the first, second, third, and fourth quartiles was 54%, 65%, 71%, and 78%, respectively. Lower baseline TIR was associated with lower TIR on CLC (r = 0.69, P < 0.001). However, lower baseline TIR was also associated with greater improvement in TIR on CLC (r = -0.81, P < 0.001). During CLC, participants in the highest versus lowest TIR-quartile administered more user-initiated boluses daily (8.5 +/- 2.8 vs. 5.8 +/- 2.6, P < 0.001) and received fewer automated boluses (3.5 +/- 1.0 vs. 6.0 +/- 1.6, P < 0.001). Participants in the lowest (vs. the highest) TIR-quartile received more insulin per body weight (1.13 +/- 0.27 vs. 0.87 +/- 0.20 U/kg/d, P = 0.008). However, in a multivariate model adjusting for baseline TIR, user-initiated boluses and insulin-per-body-weight were no longer significant. Conclusions: Higher baseline TIR is the strongest predictor of TIR on CLC in children with T1D. However, lower baseline TIR is associated with the greatest improvement in TIR. As with open-loop systems, user engagement is important for optimal glycemic control.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.4
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available