Field evaluation of a rapid antigen test (Panbio™ COVID-19 Ag Rapid Test Device) for COVID-19 diagnosis in primary healthcare centres

Journal

CLINICAL MICROBIOLOGY AND INFECTION
Volume 27, Issue 3, Pages -

Publisher

ELSEVIER SCI LTD
DOI: 10.1016/j.cmi.2020.11.004

Keywords

COVID-19; Early diagnosis; Primary healthcare centre; Rapid antigen detection test (RAD); SARS-CoV-2

Funding

  1. Roche Diagnostics (Spain)
  2. Biomerieux Inc. (France)
  3. Pfizer (Spain)
  4. Merck (Spain)

Ask authors/readers for more resources

The Panbio (TM) COVID-19 Ag Rapid Test Device performed well as a POC test for early diagnosis of COVID-19 in primary healthcare centres, with a specificity and sensitivity of 100% and 79.6% respectively. The negative predictive value of RAD was 99% and 97.9% under estimated prevalence of 5% and 10% respectively.
Objectives: To our knowledge no previous study has assessed the performance of a rapid antigen diagnostic immunoassay (RAD) conducted at the point of care (POC). We evaluated the Panbio (TM) COVID-19 Ag Rapid Test Device for diagnosis of coronavirus 2019 disease (COVID-19) in symptomatic patients (n = 412) attending primary healthcare centres. Methods: RAD was performed immediately after sampling following the manufacturer's instructions (reading at 15 min). RT-PCRs were carried out within 24 h of specimen collection. Samples displaying discordant results were processed for culture in Vero E6 cells. Presence of severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2) in cell cultures was confirmed by RT-PCR. Results: Out of 412 patients, 43 (10.4%) tested positive by RT-PCR and RAD, and 358 (86.9%) tested negative by both methods; discordant results (RT-PCR+/RAD-) were obtained in 11 patients (2.7%). Overall specificity and sensitivity of rapid antigen detection (RAD) was 100% (95%CI 98.7-100%) and 79.6% (95%CI 67.0-88.8%), respectively, taking RT-PCR as the reference. Overall RAD negative predictive value for an estimated prevalence of 5% and 10% was 99% (95%CI 97.4-99.6%) and 97.9% (95%CI 95.9 -98.9), respectively. SARS-CoV-2 could not be cultured from specimens yielding RT-PCR+/RAD- results (n = 11). Conclusion: The Panbio (TM) COVID-19 Ag Rapid Test Device performed well as a POC test for early diagnosis of COVID-19 in primary healthcare centres. More crucially, the data suggested that patients with RT-PCRproven COVID-19 testing negative by RAD are unlikely to be infectious. (C) 2020 European Society of Clinical Microbiology and Infectious Diseases. Published by Elsevier Ltd. All rights reserved.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.7
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available