4.5 Article

Limits and Opportunities of SARS-CoV-2 Antigen Rapid Tests: An Experienced-Based Perspective

Journal

PATHOGENS
Volume 10, Issue 1, Pages -

Publisher

MDPI
DOI: 10.3390/pathogens10010038

Keywords

rapid antigen tests; SARS-CoV-2; COVID-19

Categories

Ask authors/readers for more resources

This study analyzed the performance of three different rapid antigen tests in a clinical setting, showing varying specificities and sensitivities which did not entirely meet all expectations. Therefore, careful evaluation is needed when considering their benefits for specific testing strategies and settings.
Background: Due to the steadily rising case numbers of SARS-CoV-2 infections worldwide, there is an increasing need for reliable rapid diagnostic devices in addition to existing gold standard PCR methods. Actually, public attention is focused on antigen assays including lateral flow tests (LFTs) as a diagnostic alternative. Therefore, different LFTs were analyzed regarding their performance in a clinical setting. Material and Methods: A pilot sample panel of 13 bronchoalveolar fluids (BALFs) and 60 throat washing (TW) samples with confirmed PCR results, as well as eight throat washes invalid by PCR, were tested with the BIOCREDIT test (RapiGEN), the Panbio(TM) assay (Abbott), and the SARS-CoV-2 rapid antigen test (Roche). Conclusion: The analyzed antigen test showed an interassay correlation of 27.4%, with overall specificities ranging from 19.4% to 87.1%, while sensitivities of the respective tests ranged between 33.3% and 88.1%. Because these assays did not entirely meet all high expectations, their benefit has to be carefully evaluated for the respective test strategy and setting.

Authors

I am an author on this paper
Click your name to claim this paper and add it to your profile.

Reviews

Primary Rating

4.5
Not enough ratings

Secondary Ratings

Novelty
-
Significance
-
Scientific rigor
-
Rate this paper

Recommended

No Data Available
No Data Available